Omega FTB600B manual Frequency Ranges for, Cable Requirements

Page 12

4

FTB600B Series

Infrared Sensors

 

 

 

C. Frequency Ranges for

 

 

(non-cartridge models)

pulses/l

Model

Frequency (Hz)

K-Factor

FTB601B

60 to 1200

 

36,000

FTB602B

40 to 1200

 

8,000

FTB603B

27 to 800

 

3,200

FTB604B

20 to 600

 

1,200

FTB605B

18.75 to 562

 

450

FTB606B

15 to 450

 

225

D. Cable Requirements

1.20 to 22 AWG (American Wire Gauge)

2.4 conductor-shielded cable.

N O T E

Avoid influences of strong electromagnetic forces as they can damage components on the PC boards.

8

Image 12
Contents User’sGuide Servicing North America FTB600B Series Ultra-Low Flow Sensors SectionFTB600B Series Ultra-Low Flow Sensors FTB600B Series Installation IntroductionGeneral Description Principle of OperationMaterial Characteristics of Pvdf Material of constructionTemperature FTB600B Series Operating ParametersFlow Ranges FTB600B Series Recommended Viscosity Filter RecommendationsFTB600B Series Silicone Treatment FTB600B Series Infrared Sensors Supply VoltageFrequency Output Frequency Ranges for Cable RequirementsFTB600B Series Infrared Sensors Page Supply Voltage and Signal Output Connections Piping Table FTB600B Series Installation FTB600B Series Specifications FTB600B Series Pressure Drop Curves KPa Liters/minuteKPa FTB600B Series Pressure Drop Curves Pressure Drop FTB600B Series Liquid Turbine Flowmeter FTB604 160 140 120 100 180 160 140 120 100 WARRANTY/ Disclaimer Shop online at omega.com PH/CONDUCTIVITY