Omega FTB600B manual

Page 14

FTB600B Series

5 Installation

Section 5 - Installing the FTB600B Series in the fluid line.

A. Make sure the fluid is compatible with PVDF (polyvinylidene fluoride) and meets viscosity, pressure and temperature parameters of the FTB600B Series. The fluid must also meet filtration requirements as listed in 3D.

B. Install the FTB600B series in the fluid line with the arrow pointing in the direction of the flow.

C. While installing the FTB600B Series in the fluid line be careful not to over-torque the end caps (on hose-barbed flowmeters) or other fittings on the flowmeter. Due to the relatively soft composition of PVDF the body or threads can be permanently distorted.

D. Attach wires to the readout display with the display and power off. Not only will this help to avoid a potential shock hazard, but it can also help prevent an error in hooking the flow meter to an incorrect 115 Vac supply.

E. Connect digital display to power supply and enter scaling factors for both the rate and total. Follow the manufacturers instructions for programming the digital display.

F. The FTB600B Series is now ready for use.

G. See Table 5-1 for straight pipe recommendations.

10

Image 14
Contents User’sGuide Servicing North America FTB600B Series Ultra-Low Flow Sensors SectionFTB600B Series Ultra-Low Flow Sensors General Description FTB600B Series InstallationIntroduction Principle of OperationMaterial Characteristics of Pvdf Material of constructionFlow Ranges TemperatureFTB600B Series Operating Parameters FTB600B Series Recommended Viscosity Filter RecommendationsFTB600B Series Silicone Treatment Frequency Output FTB600B Series Infrared SensorsSupply Voltage Frequency Ranges for Cable RequirementsFTB600B Series Infrared Sensors Page Supply Voltage and Signal Output Connections Piping Table FTB600B Series Installation FTB600B Series Specifications FTB600B Series Pressure Drop Curves KPa Liters/minuteKPa FTB600B Series Pressure Drop Curves Pressure Drop FTB600B Series Liquid Turbine Flowmeter FTB604 160 140 120 100 180 160 140 120 100 WARRANTY/ Disclaimer Shop online at omega.com PH/CONDUCTIVITY