3.Tubing Connections

CAUTION: DO NOT FLOW ANY LIQUID THROUGH A GAS FLOW SENSOR OR FLOW METER. THIS WILL SERIOUSLY EFFECT THE PERFORMANCE AND VOID THE WARRANTY.

CAUTION: DO NOT FLOW A HIGH PRESSURE GAS THROUGH THE SENSOR. A BURST OF HIGH PRESSURE GAS MAY CAUSE THE UNIT TO BE DAMAGED. ALWAYS FLOW GAS AT A PRESSURE AND FLOW RATE WITHIN THE SPECIFCATIONS OF THE UNIT

All tubing must be clean, dry and without crimps, burrs or scratches.

Only use the factory installed fittings on the unit. If the fittings are removed the calibration of the unit may be effected and leaking may occur. If different fittings are required please contact the Omega Engineering Company Service Department for assistance.

When connecting the sensor to the tubing a wrench should be used to stop the fitting rotating in the sensor body. Take care not to over tighten the fittings or leaking may occur.

Connecting and tightening the Fittings

(FLR 1000-ST-D shown, other models similar)

Caution: DO NOT over tighten the fittings into the flow sensor or flow meter body. Excessive force may damage the sensor body. This type of damage will NOT be repaired under warranty.

The flow direction for the unit is clearly marked on the label. Do not reverse the flow direction or the unit will not function correctly.

pg. 7 of 29

Image 7
Contents FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases Page B L E O F C O N T E N T S Unpacking IntroductionProduct Overview and Principle of Operation Installation Mounting the Flow Sensor or Flow Meter General ConsiderationsTubing Connections Electrical Connections Overview Connecting a Cable Assembly Electrical Connections Voltage Output Units Using a 0-5VDC Output Power Adapter Package 5VDC Analog Outputs OperationStart Up Flow ReadingsFlows Above the Maximum Rated Flow Operating at Flow Rates Outside the Calibrated Flow RangeFlows Below the Minimum Rated Flow Corrected Flow = Flow reading X Correction Factor Zero AdjustmentsUsing Flow Sensors or Flow Meters with Different Gases RecalibrationPage Returning Units for Repair or Recalibration Maintenance and Product CareCleaning and Flushing GeneralFLR1000 FLR1000-D FLR1000-BR-D Specifications For Gas UnitsPin Cable Wire Color Description Connector Pin And Wire Color Cross ReferenceDimensions FLR 1000 For Gases 10.0 l/min Units Acetal Fittings Shown Page Page Page Page Page Symptom Possible Cause Method of Correction Troubleshooting GuideSymptom Possible Cause Method of Correction Page M4575/0807 pg