C.Operation
CAUTION: DO NOT FLOW ANY LIQUID THROUGH A GAS
FLOW SENSOR OR FLOW METER. THIS WILL SERIOUSLY
EFFECT THE PERFORMANCE AND VOID THE WARRANTY.
1.Start Up
Before applying power to the unit check all tubing and electrical connections. Once correct installation is verified switch on the power.
2.Flow Readings
The FLR 1000 Series for gases provide a
The FLR
Each unit is factory calibrated for a specific flow range. The flow range is shown on the unit’s label and calibration certificate. Units are calibrated using AIR as the reference media unless otherwise noted. Calibrations are carried out with the mounting feet horizontal on a flat horizontal surface.
a)0-5VDC Analog Outputs
By monitoring the voltage output signal it is possible to determine the flow rate of the gas. Units are configured so that an output signal of 5.0VDC is provided when the maximum flow (i.e. Full Scale flow) is passing through the unit. The output signal is linear and scaleable enabling calculation of flow rates within the sensor’s range.
For example:
For a flow range of
At 500ml/min the output signal would be 5VDC
If the output signal were 3.5VDC then the flow rate would be:
(500 ÷ 5) × 3.5 = 350ml/min
If the maximum flow rate is exceeded
b)Units With an Integral Display
The FLR
pg. 12 of 29