Step 3 Connecting the signal cables
(1)Connect the strobe timing signal cable to the PC terminal for flash syncronization on the side of the digital camera.
Rotate the screw on the connector of the cable in clockwise direction to lock the cable.
Rotate the screw in counterclockwise direction when disconnecting the cable.
(2)Connect the release signal cable to the remote port on side of the digital camera.
Pull the metal part of the connector to unlock the connector when disconnecting the cable.
6.Before Calling a Service Person
Whenever a problem occurs in the system using this adapter or when there is a mistake in operation, error will be indicated by message on the retinal camera.
In this case, take the countermeasures indicated below according to the cause.
For details and other errors, please also refer to the operation manuals of the respective instruments.
NOTE: If function is still not restored or when error display does not disappear, contact Canon repre- sentative or distributor.
Message on Retinal Camera | Cause | Remedy |
|
|
|
| Digital camera is not ON. | Turn it ON. |
|
|
|
| CF card has ran out of free | Replace with a CF card which |
FILM UNIT ERROR | space. | has enough space. |
| Signal cables between the | Disconnect the cables and re- |
| adapter and digital camera are | |
| connect them properly. | |
| not connected properly. | |
|
| |
|
|
|