Fluke 187 user manual Physical Specifications

Page 83

Model 187 & 189

Users Manual

Physical Specifications

Display (LCD)

Operating Temperature

Storage Temperature

Temperature Coefficient

Relative Humidity

Altitude

Battery Type

Battery Life

Shock Vibration

Electromagnetic Compatibility (EMC)

Size

Weight

Warranty

Calibration Interval

Digital: 50000/5000 counts primary display, 5000 counts secondary display; updates 4/second.

Analog: 51 segments, updates 40/second.

20 °C to + 55 °C

40 °C to + 60 °C

0.05 x (specified accuracy) / °C (<18 °C or >28 °C)

0 % to 90 % (0 °C to 35 °C)

0 % to 70 % (35 °C to 55 °C)

Operating: 0-2000 meters per EN61010 CAT III, 1000 V; CAT IV, 600 V 0-3000 meters per EN61010 CAT II, 1000 V; EN61010 CAT III, 600 V; CAT IV, 300 V

Storage: 10000 meters

4 AA Alkaline, NEDA 15A or LR6

72 Hours typical (with backlight off)

Per MIL-T-PRF 28800 for Class II instruments

Susceptibility and Emissions: Commercial Limits per EN61326-1

10.0cm x 20.3 cm x 5.0 cm (3.94 in x 8.00 in x 1.97 in) (Not Including Accessory Mount)

545 grams (1.2 lbs.)

Lifetime

1 year

7-2

Image 83
Contents Model 187 Lifetime Limited Warranty Table of Contents Making Measurements Model 187Iii Specifications MaintenanceList of Tables List of Figures Viii Contacting Fluke Safety InformationBefore You Start WWarning Safety InformationBefore You Start International Electrical Symbols SymbolsTurning the Meter On Getting AcquaintedIntroduction Automatic Power Off Battery ConsiderationsAutomatic Backlight Off Low Battery IndicationRotary Switch Rotary SwitchPushbuttons PushbuttonsBlue Key Function Rotary Switch Selections Position Rotary Switch FunctionView MEM Rotary Switch PositionsButton Pushbuttons Yellow Button Description Function Yellow Button Function Pushbuttons Button DescriptionPrimary Display Selecting the RangeUnderstanding the Display Secondary Display Bar GraphDisplay Features Number Feature Display Features Description 8.8.8 Display Features NumberDisplay Features Description Number FeatureFMS8888 HM8888 8.8 Display Features Number 51000Input Terminals Using the Input TerminalsDisplay Hold and AutoHOLD Using Display HoldUsing MIN MAX Using AutoHOLDModel 187 Min Max Avg Using Fast MN MXUsing Hold with MIN MAX or Fast MN MX Using Relative Mode RELMeasuring Voltage Making MeasurementsAC Voltage Measurement Measuring AC VoltageDBm Display DB Measurements in AC Volts FunctionsBoth AC and DC Voltage Measurements Measuring DC VoltageAC and DC Display AC over DC DC over ACCapacitors before measuring resistance Measuring ResistanceOFF Automatically subtract this value Testing for ContinuityUsing Conductance for High Resistance Tests For in-circuit tests, turn circuit power off Conductance Measurement Measuring Capacitance Capacitance Measurement Testing DiodesDiode Test Electrically live circuits Measuring TemperatureShock, do not connect the thermocouples to Input Ranges Measuring CurrentMeasuring Current Input Alert FeatureFollowing are some tips for measuring current Measuring AC Current11. AC Current Measurement Measuring DC Current Current through one component Total current to circuit13. Functions Allowing Frequency Measurement Measuring Frequency14. Hz Display Measuring Duty CycleMeasures Negative Pulse Measures Positive Pulse16. Duty Cycle Display Measuring Pulse WidthStart and Stop on Negative Slope Start and Stop on Positive Slope18. Pulse Width Display Logged Readings Memory Using Memory & Communications FeaturesTypes of Memory Saved Readings MemoryStopping Logging Storing Saved ReadingsStarting Logging To avoid electric shock, disconnect the test Viewing Memory DataMEM View DisplayClearing Memory Using Communications 187Selecting Setup Options Changing the Default SettingsOption Function Specific Setup SelectionsFactory Default PreconditionChoices Common Setup SelectionsAdjusting the Temperature Offset Changing the Default Settings Setting the Power Off Timeout Selecting Display Resolution 3 1/2 or 4 1/2 DigitsSetting the Line Main Frequency Setting the 24-Hour ClockSaving Setup Options Returning to Factory DefaultsTesting the Fuses MaintenanceIntroduction General Maintenance Replace F2 if Replace F1 if Displayed Replacing the Batteries Battery and Fuse Replacement Case of Difficulty Replacing the FusesUser-Replaceable Parts MP8 MP9 Safety and Compliances SpecificationsPhysical Specifications Feature Summary Basic Specifications Detailed Accuracy Specifications KHz-100 kHz Accuracy Function45-1 kHz 20 kHzDC a 500.00 Ω Capacitance Diode Test MIN MAX AVG Burden Voltage A, mA, µA Frequency Counter SensitivityInput Characteristics