SECTION 2—FEATURES

Oxygen Pressure Gauge: Enables the user to monitor the contents of the compressed oxygen cylinder and is protected by a rubber guard.

Selector Switch: Enables the user to select the desired setting, as well as “OFF” and “CF” (continuous flow). When not in use, the switch should be turned to the “OFF” position.

Oxygen Supply Outlet: Use this fitting to attach a standard cannula.

Cylinder Alignment Pins: When assembling the unit, these parallel pins MUST be inserted into the holes on the CGA870 post valve.

CGA870 Seal Washer [Metal Washer with Viton or equivalent Seal]: Creates the interface between the post valve and the conserver regulator. Besides offering a rugged interface, it also surrounds the oxygen path in a ring of stainless steel or brass.

Vent Hole: Maintains proper internal pressure. DO NOT obstruct with any object, such as a label or tight‐fitting carrying bag.

NOTE: Use only a CGA870 Seal Washer [Metal Washer with Viton Seal]. DO NOT use a plastic compression gasket.

Pneumatic Oxygen Conserver

14

Part No 1160840

Image 14
Contents Invacare Pneumatic Oxygen Conserver Accessories Warning Table of Contents Register Your Product Intended Use Special NotesReceiving Inspection Disposal Instructions Principles of OperationSpecial Notes Introduction General Guidelines Operating InformationGeneral Guidelines General Guidelines General Guidelines Features Model IOC100PFeatures Connection view of the IOC100P with CGA870 Installing the Conserving Regulator Corresponding holes on Operating Instructions Cannula Placement Operating Instructions Selector Switch Removing the Conserving Regulator Typical Product Parameters RegulatorStorage Oxygen Cylinder Duration Oxygen Duration TableMaintenance Troubleshooting Replacement Parts AccessoriesPneumatic Oxygen Conserver Pneumatic Oxygen Conserver Limited Warranty Limited Warranty USA