Invacare IOC100P manual Corresponding holes on

Page 17

SECTION 3—INSTALLING THE CONSERVING REGULATOR

NOTE: For information on how to operate the cylinder, consult your equipment provider.

NOTE: For this procedure, refer to FIGURE 3.1.

1.Loosen the cylinder post valve T‐Handle.

2.Lower the conserver over any CGA870 post valve cylinder with the alignment pins toward the holes on the cylinder neck.

3.Align the two pins and

gasket with the

FIGURE 3.1 Attaching the

corresponding holes on

IOC100P to the Cylinder

the cylinder post valve.

 

NOTE: Cylinder post valve T‐Handle should be aligned with indentation on post valve.

4.While holding the unit in place, tighten the cylinder post valve by turning clockwise.

5.Attach a standard cannula [7 ft. (2.13 m) or less in length] to the oxygen supply outlet fitting.

NOTE: Tighten only by hand. The use of a tool to tighten the T‐Handle may damage the unit.

Part No 1160840

17

Pneumatic Oxygen Conserver

Image 17
Contents Invacare Pneumatic Oxygen Conserver Accessories Warning Table of Contents Register Your Product Intended Use Special NotesReceiving Inspection Principles of Operation Disposal InstructionsSpecial Notes Introduction Operating Information General GuidelinesGeneral Guidelines General Guidelines General Guidelines Model IOC100P Features Features Connection view of the IOC100P with CGA870 Installing the Conserving Regulator Corresponding holes on Operating Instructions Cannula Placement Operating Instructions Selector Switch Removing the Conserving Regulator Regulator Typical Product ParametersStorage Oxygen Duration Table Oxygen Cylinder DurationMaintenance Replacement Parts Accessories TroubleshootingPneumatic Oxygen Conserver Pneumatic Oxygen Conserver Limited Warranty Limited Warranty USA
Related manuals
Manual 24 pages 50.35 Kb