Invacare IOC100P manual Troubleshooting, Replacement Parts Accessories

Page 27

SECTION 8—TROUBLESHOOTING

SECTION 8—TROUBLESHOOTING

If oxygen conserving regulator fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your Dealer/Home Care equipment provider.

PROBLEM

POSSIBLE CAUSE

 

SOLUTION

 

 

 

 

 

No flow. Unit does not

1.

Cylinder valve is

1.

Turn the cylinder

pulse.

 

closed.

 

valve to the “ON”

 

 

 

 

position.

 

2.

Regulator in “Off”

2.

Set to prescribed

 

 

position.

 

setting.

 

3.

Cylinder empty.

3.

Replace cylinder.

 

4.

Oxygen conserving

4.

Check position of

 

 

regulator not

 

cannula in nose. DO

 

 

sensing breath.

 

NOT breathe

 

 

 

 

through mouth.

 

5.

Oxygen cannula is

5.

Remove kinks. Clean

 

 

blocked or kinked.

 

or replace, if

 

 

 

 

necessary.

 

 

 

 

 

Unit flows or pulses

1.

Unit is set to the

1.

Turn the selector

continuously.

 

“CF” position.

 

switch to the

 

 

 

 

appropriate delivery

 

 

 

 

setting.

 

2.

Unit was not set to

2.

Turn the selector

 

 

“OFF” prior to

 

switch to “OFF,” wait

 

 

opening the

 

a few moments, then

 

 

cylinder valve.

 

set at proper delivery

 

 

 

 

setting.

 

3.

Vent hole is

3.

Remove obstructions,

 

 

obstructed.

 

such as labels or

 

 

 

 

carrying bag, and

 

 

 

 

resume use as usual.

 

 

 

 

 

Replacement Parts - Accessories

HF2POST6BAG Carrying Bag or HF2POST9BAG Carrying Bag

Part No 1160840

27

Pneumatic Oxygen Conserver

Image 27
Contents Invacare Pneumatic Oxygen Conserver Accessories Warning Table of Contents Register Your Product Special Notes Receiving InspectionIntended Use Principles of Operation Disposal InstructionsSpecial Notes Introduction Operating Information General GuidelinesGeneral Guidelines General Guidelines General Guidelines Model IOC100P FeaturesFeatures Connection view of the IOC100P with CGA870 Installing the Conserving Regulator Corresponding holes on Operating Instructions Cannula Placement Operating Instructions Selector Switch Removing the Conserving Regulator Regulator Typical Product ParametersStorage Oxygen Duration Table Oxygen Cylinder DurationMaintenance Replacement Parts Accessories TroubleshootingPneumatic Oxygen Conserver Pneumatic Oxygen Conserver Limited Warranty Limited Warranty USA
Related manuals
Manual 24 pages 50.35 Kb