SECTION 8—TROUBLESHOOTING
If oxygen conserving regulator fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your Dealer/Home Care equipment provider.
PROBLEM | POSSIBLE CAUSE |
| SOLUTION | |
|
|
|
|
|
No flow. Unit does not | 1. | Cylinder valve is | 1. | Turn the cylinder |
pulse. |
| closed. |
| valve to the “ON” |
|
|
|
| position. |
| 2. | Regulator in “Off” | 2. | Set to prescribed |
|
| position. |
| setting. |
| 3. | Cylinder empty. | 3. | Replace cylinder. |
| 4. | Oxygen conserving | 4. | Check position of |
|
| regulator not |
| cannula in nose. DO |
|
| sensing breath. |
| NOT breathe |
|
|
|
| through mouth. |
| 5. | Oxygen cannula is | 5. | Remove kinks. Clean |
|
| blocked or kinked. |
| or replace, if |
|
|
|
| necessary. |
|
|
|
|
|
Unit flows or pulses | 1. | Unit is set to the | 1. | Turn the selector |
continuously. |
| “CF” position. |
| switch to the |
|
|
|
| appropriate delivery |
|
|
|
| setting. |
| 2. | Unit was not set to | 2. | Turn the selector |
|
| “OFF” prior to |
| switch to “OFF,” wait |
|
| opening the |
| a few moments, then |
|
| cylinder valve. |
| set at proper delivery |
|
|
|
| setting. |
| 3. | Vent hole is | 3. | Remove obstructions, |
|
| obstructed. |
| such as labels or |
|
|
|
| carrying bag, and |
|
|
|
| resume use as usual. |
|
|
|
|
|
Replacement Parts - Accessories
HF2POST6BAG Carrying Bag or HF2POST9BAG Carrying Bag
Part No 1160840 | 27 | Pneumatic Oxygen Conserver |