Agilent Technologies 85054-90049 manual Calibration Kit Overview, Kit Contents

Page 6

General Information

Calibration Kit Overview

Calibration Kit Overview

The Agilent 85054B type-N calibration kit is used to calibrate Agilent network analyzers up to 18 GHz for measurements of components with 50Ω type-N connectors.

The standards in this calibration kit allow you to perform simple 1- or 2-port and TRM (thru–reflect–match) calibrations.

This manual describes the 85054B calibration kit and provides replacement part numbers, specifications, and procedures for using, maintaining, and troubleshooting the kit.

Kit Contents

The 85054B calibration kit contains the following:

offset opens and shorts, lowband and sliding load terminations

four type-N to 7 mm adapters

two type-N to type-N adapters

a type-N connector gage set

a 3/4 inch, 135 N-cm (12 in-lb) torque wrench for use on the type-N connectors

a spanner wrench

a data disk that contains the calibration definitions of the devices in the kit for 8510 systems and the 872x series

a data disk that contains the calibration definitions of the devices in the kit for the PNA series

Refer to Table 6-1and Figure 6-1for a complete list of kit contents and their associated part numbers.

Offset Opens and Shorts

The offset opens and shorts are built from parts that are machined to the current state-of-the-art in precision machining.

The offset short’s inner conductors have a one-piece construction, common with the shorting plane. The construction provides for extremely repeatable connections.

The offset opens have inner conductors that are supported by a strong, low-dielectric constant plastic to minimize compensation values.

Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. The lengths of the offsets in the opens and shorts are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.

1-2

85054B

Image 6
Contents User’s and Service Guide Manufacturing Part Number Print Date JuneDocumentation Warranty Contents Replaceable Parts General Information Kit Contents Calibration Kit OverviewCalibration Definitions Equipment Required but Not SuppliedIncoming Inspection Serial Numbers Recording the Device Serial NumbersSerial Number Record for the 85054B Calibration Kits Documented in This Manual Calibration Kit History85054B Kits with Serial Prefix 2906A Precision Slotless Connectors Clarifying the Sex of a Connector Preventive MaintenanceSpecifications Temperature-What to Watch Out For Environmental RequirementsEnvironmental Requirements Mechanical Characteristics Pin DepthPin Depth Limits Electrical Specifications Electrical SpecificationsCertification Electrical Specifications 85054B Use, Maintenance, and Care of the Devices ESD Protection Setup Electrostatic DischargeLook for Obvious Defects and Damage First What Causes Connector Wear?Visual Inspection Inspect the Mating Plane SurfacesCleaning Connectors Inspect the Precision Slotless Connectors femaleUse Compressed Air or Nitrogen Clean the Mating Plane Surfaces InspectGaging Connectors Connector Gage AccuracyWhen to Gage Connectors Reading the Connector GageGaging Procedures Gaging Male Type-N ConnectorsGaging Male Type-N Connectors Gaging Female Type-N Connectors Gaging Female Type-N Connectors Gaging the Sliding Load Adjusting the Sliding Load Pin Depth Gaging the Sliding Load Pin DepthGaging Connectors Adjusting the Sliding Load Pin Depth Connections How to Make a ConnectionFinal Connection Using a Torque Wrench Preliminary ConnectionWrench Positions Connecting the Sliding Load How to Separate a Connection Connecting the Sliding LoadUsing the Sliding Load 10 Using the Sliding Load Handling and Storage Performance Verification Introduction How Agilent Verifies the Devices in This KitWhere to Send a Kit for Recertification RecertificationHow Often to Recertify Recertification 85054B Troubleshooting Troubleshooting Process Troubleshooting FlowchartContacting Agilent Where to Look for More InformationReturning a Kit or Device to Agilent Contacting AgilentReturning a Kit or Device to Agilent 85054B Replaceable Parts Replaceable Parts for the 85054B Calibration Kit AdaptersItem No Description Replaceable Parts-Items Not Included in the Calibration Kit ESD Protective DevicesReplaceable Parts for the 85054B Calibration Kit More Replaceable Parts for the 85054B Calibration Kit Standard Definitions Version Changes Standard Class Assignments Class Standard Class LabelCalibration Kit Label N 50 Ω If you are performing a TRL calibration If you are performing a TRM calibrationIf you are performing an LRM calibration Standard Class Assignments Blank Forms Table A-4 Blank Form for the 8510 Network AnalyzerS11A S11B S11C S22A S22B S22C Table A-6 Blank Form for the PNA Series Network Analyzers Standard Class Assignments Setting the System Impedance Nominal Standard DefinitionsTable A-7 Standard Definitions for the 8510 Network Analyzer Calibration Kit Label N 50 Ω Frequency Standardb Offset GHzc Table A-10 Blank Form for the 8510 Network Analyzer Coax or Waveguide Standard LabelSystem Z 0 a = 50.0 Ω Frequency Standard b Offset GHz c Nominal Standard Definitions 85054B Index Index-2 85054B Index-3

85054-90049 specifications

Agilent Technologies 85054-90049 is a high-performance calibration kit designed for vector network analysis across a wide range of frequencies. This kit is an indispensable tool for engineers and technicians involved in the measurement and characterization of RF components, ensuring accuracy and reliability in their testing processes.

One of the main features of the 85054-90049 calibration kit is its ability to provide accurate calibration across multiple frequency bands, including microwave and millimeter-wave frequencies. This versatility is crucial for professionals who work with different types of RF devices, such as amplifiers, filters, and antennas. The kit supports a frequency range of up to 67 GHz, making it suitable for a variety of applications in the telecommunications and aerospace industries.

The calibration kit utilizes precision standards and adapters that ensure low insertion loss and minimal reflection. These components are engineered to provide high stability and repeatability in measurements, which is vital for maintaining quality control in manufacturing processes or during research and development stages. The accuracy of the calibration is further enhanced by the materials used in the construction of the connectors and adapters, which are designed to minimize signal degradation.

Another key technology employed in the Agilent 85054-90049 is its rugged design, intended to withstand the demands of laboratory environments as well as field applications. The kit is portable, allowing engineers to perform precision measurements wherever needed without compromising on performance. Its compact design and durable case make it an ideal choice for field applications.

The 85054-90049 calibration kit is also compatible with Agilent's range of vector network analyzers, creating a seamless user experience. The calibration process, supported by comprehensive software tools, is designed to be user-friendly, allowing technicians to perform calibrations quickly and accurately. In addition, the kit comes with detailed documentation that includes calibration procedures, ensuring that even those new to RF testing can follow along with ease.

In summary, the Agilent Technologies 85054-90049 is a state-of-the-art calibration kit that combines high performance, compatibility, and durability. Its advanced features cater to various RF testing needs, making it a vital resource for professionals in industries where precision measurements are of utmost importance. Whether in laboratories or field settings, this calibration kit stands out as a benchmark for quality in RF and microwave measurement applications.