2.5 Probe Compensation

Before taking measurements with the USBscope50, the attached x10 scope probe must be compensated to ensure that it is matched to the scope input stage8.

This ensures that the probe+scope act together to keep the input frequency response as flat as possible. Using a scope probe that has not been adjusted to match the USBscope50 can lead to inaccurate readings for higher frequency signals.

The USBscope50 includes a probe compensation output9.

This must be turned on using the USBscope50 software. Once enabled, the compensation output is a 0 to 3V (approx) square wave with 50% duty cycle and 1KHz frequency. Note that the compensation output will switch off automatically when you close the channel’s tab10.

The USBscope50 can work with scope probes that list an adjustment range that includes 16pF. It is best to use probes that have a range that puts the 16pF point near the middle of the range, for example the HP10071A type probe lists an adjustment range of 9-17pF and our tests show that whilst it is easy to compensate this probe against the USBscope50, the trimmer in the probe body is certainly quite close to the top end of its range. There are many probe choices on the market, and lots of them list ranges that span 16pF.11

To compensate your probe follow these steps:

1.Connect your x10 probe to the USBscope50 software and run the USBscope50 software

8This does not apply to x1 scope probes

9Later models have a dual pad compensation point that includes a GND pad

10This is to avoid the fast edges of the square wave possibly coupling into your measurement

11Some very high frequency low capacitance probes may not be suitable for use with the USBscope50

Elan Digital Systems Ltd.

15

USBscope50 USER’S GUIDE Iss5

Image 15
Contents USER’S Guide Contents Hardware Specification Page Overview General Figure below shows a simplified diagram of the USBscope50 Using More Than One USBscope50 This is a safety critical point so please note it carefully Disassembling The Stack USB Connections For Stacked Configurations Repeat this for each channel in the stack to identify them Stacking Do’s And Don’ts Input Ranges Probe Compensation USBscope50 USER’S Guide Iss5 This Acquisition Modes Random Interleaved Sampling USBscope50 USER’S Guide Iss5 Trigger Modes Normal Triggering Trigger Position USBscope50 USER’S Guide Iss5 10Trigger Threshold USBscope50 USER’S Guide Iss5 12FFT Plot Spectrum Analyser 13Math Functions 14Data Export Data for CH FFT for CH Before plugging in the hardware USBscope50 USER’S Guide Iss5 Windows 2K and XP17 Page Power Requirements Performance Static Performance Dynamic Performance