Chapter 2 Service

About the HP 53131A/132A Calibration Menu

About the HP 53131A/132A Calibration Menu

If you are familiar with the information provided in this section, proceed to the section titled “The HP 53131A/132A Calibration Procedures” on page 2-15 to calibrate your Counter. This section, “About the

HP 53131A/132A Calibration Menu,” can be used as a reference.

With the exception of the standard timebase oscillator, no adjustments are required for the HP 53131A/132A Counter.

Overview of the HP 53131A/132A Calibration Menu

The Calibration menu is accessed by holding the Scale & Offset key and cycling the POWER key. All of the calibration factors and security settings are stored in non-volatilememory, and do not change when power has been off or after a remote interface reset.

A calibration security code feature allows you to enter a security code (electronic key) to prevent accidental or unauthorized calibrations of the Counter. The security code is set to 53131 or 53132 (corresponding to the model) when the Counter is shipped from the factory. If you forget your security code, you can reset the security code to the model-number default by resetting all of the non-volatile memory to a default state (Contact HP for this confidential procedure).

This menu does not exist in early versions of the Counter. In the early versions (firmware revisions 3317, 3335, and 3402) of the HP 53131A/132A, the CAL: menu item resides in the Utility menu, and there is no calibration security capability. See Chapter 6, “Backdating,” in this guide for the correct calibration procedures.

2-6

Assembly-Level Service Guide

Image 102
Contents HP 53131A/132A/181A 225 MHz Counter Page Assembly-Level Service Guide Safety Considerations contd Contents Test 2 Trigger Level HP 53131A/132A Only Test 5 Option 030/050/124 Channel 3 Frequency Test 2 Channel 1 Frequency Sensitivity HP 53181A Service About the HP 53181A Calibration Menu Replacing Assemblies Retrofitting Options Index Contents Xii Assembly-Level Service Guide How to Use This Guide Repair StrategyInstrument Identification Assembly-Level Service Guide Organization Instruments Covered by this GuideHow to Order Guides Description of the 225 MHz Counters Assembly-Level Service Guide Xvii Options HardwareAccessories Available Accessories Supplied and AvailableAccessories Supplied SupportPreface Performance Tests Introduction Operational VerificationIntroduction Recommended Calibration Cycle Complete Performance TestsOptional HP-IB Verification Test RecordEquipment Required Equipment RequiredRecommended Test Equipment Power-On Self Tests HP 53131A/132A/181A Operational VerificationHP 53131A/132A/181A Operational Verification HP 53131A/132A Counter1B. Operational Verification Setup for HP 53181A HP 53131A/132A/181A Operational Verification HP 53181ACounter HP 3458A Run Self TestTermination Check Digital MultimeterPress Channel 1 50Ω/1MΩ key Counter Setup External Arm Test HP 53131A/132A OnlyEquipment HP 8130A Pulse Generator ProcedureSynthesized Signal Generator HP 8663AExternal Timebase Tests MHz External Timebase Input HP 53131A and HP 53181A Only HP 3325B Synthesizer #1HP 3325B Synthesizer #2 Counter MHz External Timebase Input HP 53131A and HP 53181A Only MHz External Timebase Input for the HP 53132A HP 53131A/132A Complete Performance Tests HP 53131A/132A Complete Performance TestsHP 53131A/132A Performance Tests Equipment Counter SetupTest 1 Time Interval HP 53131A/132A Only Test 1 Time Interval HP 53131A/132A OnlyProcedure HP 8130A HP53131A/132Pulse Generator CounterTest 2 Trigger Level HP 53131A/132A Only Test 2 Trigger Level HP 53131A/132A OnlyTrigger Level Test Setup HP 3325B HP 53131A/132A Synthesizer CounterThis is the trigger level it should be 0.0 V ±15 mV Equipment HP 53131A/132A Ref. Counter HP 8663A Synthesized Signal GeneratorKHz to 100 MHz Sensitivity for HP 53131A/132A HP 8663A SynthesizedPerformance Tests MHz to 200 MHz Sensitivity for HP 53131A/132A MHz to 225 MHz Sensitivity for HP 53131A/132A Equipment Channels 1 and 2 Frequency Accuracy Test Setup HP 53131A/132A CounterPerformance Tests Equipment HP 8663A Synthesized Signal Generator HP 53131A/132A CounterSweep Oscillator To 3.0 GHz Sensitivity for Option 030 only Sweep the frequencies from 200 MHz to 5.0 GHz Sweep the frequencies from 200 MHz to 12.4 GHz Gate & ExtArm, Time Time 1.000 s On the Counter, press Run key Test 7 Peak Volts, Channels 1 and 2 HP 53131A/132A Only Test 7 Peak Volts, Channels 1 and 2 HP 53131A/132A OnlyVolt Peak HP 3325B HP 53131A/132ASynthesizer CounterVolt Peak HP 53131A/132A HP-IB Verification Program Optional HP 53131A/132A HP-IB Verification Program OptionalHP 53131A/132A Performance Test Record Page 1 HP 53131A/132A Performance Test Record Page 1HP 53131A/132A Performance Test Record Page 2 HP 53131A/132A Performance Test Record Page 2HP 53131A/132A Performance Test Record Page 3 HP 53131A/132A Performance Test Record Page 3HP 53131A/132A Performance Test Record Page 4 HP 53131A/132A Performance Test Record Page 4HP 53131A/132A Performance Test Record Page 5 HP 53131A/132A Performance Test Record Page 5HP 53131A/132A Performance Test Record Page 6 HP 53131A/132A Performance Test Record Page 6HP 53181A Complete Performance Tests HP 53181A Complete Performance TestsHP 53181A Performance Tests Test 1 Trigger Level HP 53181A Only Test 1 Trigger Level HP 53181A Only11. Trigger Level Test Setup HP 3325B HP 53181ASynthesizer CounterThis is the trigger level it should be 0.0 V ±15 mV Test 2 Channel 1 Frequency Sensitivity HP 53181A Only Test 2 Channel 1 Frequency Sensitivity HP 53181A Only12. Channels 1 Frequency Sensitivity Test Setup KHz to 100 MHz Sensitivity for HP 53181ATest 2 Channel 1 Frequency Sensitivity HP 53181A Only MHz to 200 MHz Sensitivity for HP 53181A MHz to 225 MHz Sensitivity for HP 53181ATest 3 Channel 1 Frequency Accuracy HP 53181A Only Test 3 Channel 1 Frequency Accuracy HP 53181A OnlyHP 8663A Synthesized Signal Generator HP 53181A CounterSweep Oscillator Test 3 Channel 1 Frequency Accuracy HP 53181A Only Equipment MHz to 1.5 GHz Sensitivity for Option 015 Only 14. Option 015/030/050 Channel 2 Frequency TestMHz to 2.7 GHz Sensitivity for Option 030 Only MHz to 5.0 GHz Sensitivity for Option 050 only MHz to 12.4 GHz Sensitivity for Option 124 only MHz to 1.5 GHz Accuracy Test for Option Connect equipment as shown in Figure Minimum to Maximum Accuracy Test for Options 030/050/124 Verify the HP 53181A Counter reads 100 MHz ±70 mHz Performance Tests Test 6 Peak Volts, Channel 1 HP 53181A Only Test 6 Peak Volts, Channel 1 HP 53181A Only16. Peak Volts Test Setup HP 53181A HP-IB Verification Program Optional HP 53181A HP-IB Verification Program OptionalHP 53181A Performance Test Record Page 1 HP 53181A Performance Test Record Page 1HP 53181A Performance Test Record Page 2 HP 53181A Performance Test Record Page 2HP 53181A Performance Test Record Page 3 HP 53181A Performance Test Record Page 3HP 53181A Performance Test Record Page 4 HP 53181A Performance Test Record Page 4Service Service Returning the Instrument to Hewlett-Packard for Service Returning the Instrument to Hewlett-Packard for ServiceTo Provide Repair Information To Pack in the Original Packaging Materials To Pack in the Commercially Available Materials About the HP 53131A/132A Calibration Menu Overview of the HP 53131A/132A Calibration MenuAbout the HP 53131A/132A Calibration Menu CAL Unsecure Calibration procedures HP 53131A/132A Calibration Menu TreeTo View the Calibration Menu and Security Status To Unsecure for CalibrationTo Initiate the Calibration Routines Information about Calti QUIK? and Calti FINE? About the HP 53131A/132A Calibration Menu HP 8130A HP 53131A/132A Pulse Generator Counter HP 59992A J06 Time Interval CalibratorTo Secure Against Calibration To Change to a New Security CodeTo View the Calibration Count To Get Help with the Calibration MenuFirst Determine the Counter Firmware Revision HP 53131A/132A Calibration ProceduresHP 53131A/132A Calibration Procedures To Calibrate the Offset for Channels 1BAD Code HP 53131A/132A Calibration Procedures DC Power Supply HP 53131A/132A Counter To Calibrate the Gain for Channels 1Digital Multimeter To Calibrate Time Interval Press Enter key againCAL TI QUIK? Calibration HP 8130A HP53131A/132A Pulse Generator CounterCAL TI FINE? Calibration HP 8130A Pulse GeneratorTI CAL Pass To Calibrate the Standard Timebase House Standard MHz Reference HP 53131A/132A CounterTo Calibrate the High Stability Timebase Option Medium, High, or Ultra-HighAbout the HP 53181A Calibration Menu Overview of the HP 53181A Calibration MenuAbout the HP 53181A Calibration Menu About the HP 53181A Calibration Menu Key to access this menu HP 53181A Calibration Menu TreeTo View the Calibration Menu and Security Status To Initiate the Calibration Routines To Change to New the Security Code HP 53181A Calibration Procedures HP 53181A Calibration ProceduresTo Calibrate the Offset for Channel BAD Code HP 53181A DC Power Supply Counter Digital Multimeter To Calibrate the Gain for ChannelHouse Standard MHz Reference HP 53181A Counter To Calibrate the High Stability Timebase Option Pre-Troubleshooting Information Safety ConsiderationsPre-Troubleshooting Information Recommended Test Equipment Interruption is ProhibitedElectrostatic Discharge Disassembly and Reassembly SpecificsRepair Considerations Surface Mount RepairAfter Service Considerations Product Safety ChecksProduct Performance Checks Assembly Identification and Location HP 53131A Assembly Identification53132-60014 53131-69014 Assembly Name HP Part No Exchange Part A1 Motherboard Series 8A. Instrument Top Internal View 8B. Instrument Top Internal View Option Power Supply Check Troubleshooting the CounterMotherboard Assembly A4 AC Power Supply Voltages Wire Color Voltage ToleranceOverview of the Self-Test Routines Test Diagnosing the Faulty Assembly by Using Self TestsTo Run the Test-All Self Test To Run the Individual Self Tests Test Loop to OFF11B. HP 53181A Self Test Troubleshooting Setup Troubleshooting Failure Indicator Assembly at FaultReplacing Assemblies Replacing Assemblies Tools Required Do This FirstTools Required To Remove the Cover Rear Bezel RemovalBottom View for Cover Removal To Remove the Front Bezel To Remove the Front BezelFront Bezel Removal To Remove the Front Bezel To Remove A1 Motherboard Assembly To Remove A1 Motherboard AssemblyA1 Motherboard Assembly Removal To Remove the A2 Display Board, Keypads, and Window To Remove the A2 Display Board, Keypads, and WindowTo Remove the A2 Display Board, Keypads, and Window To Remove A3 1.5/3.0/5.0/12.4 GHz Channel Assembly Option Replacing Assemblies015/030/050/124 Optional Channel Input Assembly Removal To Remove A4 AC Power Supply Assembly To Remove A4 AC Power Supply Assembly10. Location of the Rubber Foot To Remove A5 DC Power Input Assembly Option To Remove A5 DC Power Input Assembly Option11. DC Power Input Assembly Option 002 Removal Remove the cover Remove A6 High Stability Timebase Assembly To Remove the Rear Terminals Option To Remove the Rear Terminals Option13. Rear Terminals Option 060 Removal To Remove the Rear Terminals Option To Remove the Rear Terminals Option Retrofitting Options Retrofitting Options Retrofitting Options Preliminary Procedure To Retrofit A5 DC Power Input Assembly OptionOption 002 DC Power Input Assembly Parts To Retrofit A5 DC Power Input Assembly OptionRetrofitting Procedure 1A. DC Power Input Assembly Retrofitting1B. DC Power Input Assembly Retrofitting To Retrofit A5 DC Power Input Assembly Option 1C. DC Power Input Assembly Retrofitting 1D. DC Power Input Assembly Retrofitting Dc-supply Option 001 Medium Stability Timebase Assembly Parts Option 010 High Stability Timebase Assembly PartsOption 012 Ultra-High Stability Timebase Assembly Parts Turn off the Counter and remove power cord Remove the cover High Stability Timebase Retrofitting Option 001, 010 To Retrofit the Rear Terminals Option Option 060 Rear Terminals PartsTo Retrofit the Rear Terminals Option 3A. Plug Removal for Rear Terminals Retrofitting 3B. Rear Terminals Retrofitting To the rear of the chassis To Retrofit the Rear Terminals Option Replaceable Parts Exchange Assemblies Exchange AssembliesReference Designations Replaceable PartsReference Designations How To Order a Part How To Order a PartParts Identification Contacting Hewlett-Packard By telephoneFor mail correspondence, use the address below Cabinet Parts and Hardware Cabinet Parts and HardwareReference Designation Mfr Mfr Part Number CodeOption MP2 MP3 Manufacturer’s Code List Mfr Code Manufacturer Name Address Zip CodeBackdating Manual Changes Older InstrumentsInstrument this Guide Directly Applies To Change 1 Below Series Prefix 3711A Change 2 New A3 5/12.4 GHz Prescaler AssemblyBackdating Hardware Backdating Hardware10. A4 AC Power Supply -9V Output Check Option 010/012 Only Change 3 New Tcxo boardsChange 4 Series 3710A to 3705A Change 5 Series 3705A to 3546A, 3548A Change 6 Below Series Prefix 3416AChange 7 Below Serial Number 3349A01906 Change 8 Below Serial Prefix 3404AChange 1 Firmware Revisions 3317, 3335, 3402 Backdating FirmwareAbout the HP 53131A Calibration Menu Backdating FirmwareHP 53131A Utility/Calibration Menu Tree About the HP 53131A Calibration MenuHP 53131A Calibration Procedures HP 53131A Calibration ProceduresDC Power Supply HP 53131A Counter To Calibrate Time Interval Equipment HP 8130A HP53131APulse Generator CounterHouse Standard MHz Reference HP 53131A Counter Medium or High Change 1 53131A, Serial Prefix 3710A and Below Backdating SpecificationsBackdating Specifications Change 2 53132A, Serial Prefix 3546A and BelowChange 3 53181A, Serial Prefix 3711A02841 and Below Backdating Specifications HP 53131A/132A Specifications Channel 1 & 2 Input Specifications1 Channel 1 & 2 Input Characteristics1Instrument Inputs External Arm Input Specifications Channel 3 Input Specifications4Channel 3 Input Characteristics External Arm Input CharacteristicsTime Base Time Base Output SpecificationsTime Base Internal Time Base StabilityMeasurement Specifications Measurement SpecificationsFrequency, Period HP 53131A-Worst Case RMS Resolution Automatic or External ArmingHP 53132A-Worst Case RMS Resolution Frequency Measurement Example × FrequencyTime Interval LSDPhase Pulse WidthRise/Fall Time Duty Cycle TotalizePeak Volts Measurement Definitions Definitions of Systematic Uncertainty Terms Trigger ErrorTrigger Level Timing Error Measurement DefinitionsTypical Versus Worst Case Specifications Time Interval Measurement LSD DefinitionAdditional Considerations for Digits and Time Arming Modes Gate Time Measurement Throughput Measurement Arming and ProcessingMeasurement Arming and Processing Measurement ArmingMeasurement Statistics Measurement LimitsGeneral Information General InformationHP 53181A Specifications Channel 1 Input Specifications Channel 1 Input CharacteristicsChannel 2 Input Specifications3 Channel 2 Input CharacteristicsTime Base Where N = 1 for Ch1 Frequency 1 MHz Worst Case RMS Resolution Frequency Measurement Example Ratio 1 ⁄ Ch2 Freq × Gate Time Ratio 2 ⁄ Typical Versus Worst Case Specifications Available Statistics Power is removed from the counter General Information Index Index-2 Assembly-Level Service Guide Assembly-Level Service Guide Index-3 Index-4 Assembly-Level Service Guide Contacting Hewlett-Packard Page From front matter Manual Part Number