Chapter 3 Replacing Assemblies

 

 

 

To Remove A3 1.5/3.0/5.0/12.4 GHz Channel Assembly (Option

 

 

015/030/050/124)

 

 

 

 

 

 

 

To Remove A3 1.5/3.0/5.0/12.4 GHz Channel

 

 

 

Assembly (Option 015/030/050/124)

 

 

 

 

NOTE

 

The 1.5 GHz Channel 2 Assembly is available only for the HP 53181A

 

 

 

Frequency Counter.

 

 

 

 

1Remove the cover.

See the section titled “To Remove the Cover” in this chapter.

2Using the 14-mm deep-socket spin tight, remove the BNC nut (H3) from the optional Channel 3 (or Channel 2 for HP 53181A) input connector (Options 015/030 only). For Option 050 or 124 remove the knurled nut using a pair of pliers.

3Pull W1, which is the optional channel assembly cable, out of the Channel 3 (or 2) opening in the front panel assembly as shown in Figure 3-8. For the Option 050 or 124, the connector is part of the Option assembly.

CAUTION

When reassembling A3 (Option 015/030) Assembly, ensure that cable W1

 

is not placed or folded on top of the input circuits of A1 Motherboard

 

Assembly or is tucked behind A3 as the Counter may start reading or

 

measuring noise.

 

 

3

Assembly-Level Service Guide

3-13

Image 161
Contents HP 53131A/132A/181A 225 MHz Counter Page Assembly-Level Service Guide Safety Considerations contd Contents Test 2 Trigger Level HP 53131A/132A Only Test 5 Option 030/050/124 Channel 3 Frequency Test 2 Channel 1 Frequency Sensitivity HP 53181A Service About the HP 53181A Calibration Menu Replacing Assemblies Retrofitting Options Index Contents Xii Assembly-Level Service Guide Instrument Identification How to Use This GuideRepair Strategy Instruments Covered by this Guide Assembly-Level Service Guide OrganizationHow to Order Guides Description of the 225 MHz Counters Assembly-Level Service Guide Xvii Hardware OptionsAccessories Supplied Accessories Supplied and AvailableAccessories Available SupportPreface Performance Tests Introduction IntroductionOperational Verification Optional HP-IB Verification Complete Performance TestsRecommended Calibration Cycle Test RecordRecommended Test Equipment Equipment RequiredEquipment Required HP 53131A/132A/181A Operational Verification Power-On Self TestsHP 53131A/132A/181A Operational Verification Counter HP 53131A/132AHP 53131A/132A/181A Operational Verification HP 53181A 1B. Operational Verification Setup for HP 53181ATermination Check Run Self TestCounter HP 3458A Digital MultimeterPress Channel 1 50Ω/1MΩ key Equipment Counter SetupExternal Arm Test HP 53131A/132A Only Synthesized Signal Generator ProcedureHP 8130A Pulse Generator HP 8663AExternal Timebase Tests HP 3325B Synthesizer #2 Counter MHz External Timebase Input HP 53131A and HP 53181A OnlyHP 3325B Synthesizer #1 MHz External Timebase Input HP 53131A and HP 53181A Only MHz External Timebase Input for the HP 53132A HP 53131A/132A Performance Tests HP 53131A/132A Complete Performance TestsHP 53131A/132A Complete Performance Tests Test 1 Time Interval HP 53131A/132A Only Counter SetupEquipment Test 1 Time Interval HP 53131A/132A OnlyHP 8130A HP53131A/132Pulse Generator Counter ProcedureTest 2 Trigger Level HP 53131A/132A Only Test 2 Trigger Level HP 53131A/132A OnlyHP 3325B HP 53131A/132A Synthesizer Counter Trigger Level Test SetupThis is the trigger level it should be 0.0 V ±15 mV Equipment KHz to 100 MHz Sensitivity for HP 53131A/132A Counter HP 8663A Synthesized Signal GeneratorHP 53131A/132A Ref. HP 8663A SynthesizedPerformance Tests MHz to 200 MHz Sensitivity for HP 53131A/132A MHz to 225 MHz Sensitivity for HP 53131A/132A Equipment HP 53131A/132A Counter Channels 1 and 2 Frequency Accuracy Test SetupPerformance Tests Equipment Sweep Oscillator HP 8663A Synthesized Signal GeneratorHP 53131A/132A Counter To 3.0 GHz Sensitivity for Option 030 only Sweep the frequencies from 200 MHz to 5.0 GHz Sweep the frequencies from 200 MHz to 12.4 GHz Gate & ExtArm, Time Time 1.000 s On the Counter, press Run key Test 7 Peak Volts, Channels 1 and 2 HP 53131A/132A Only Test 7 Peak Volts, Channels 1 and 2 HP 53131A/132A OnlyHP 3325B HP 53131A/132ASynthesizer Counter Volt PeakVolt Peak HP 53131A/132A HP-IB Verification Program Optional HP 53131A/132A HP-IB Verification Program OptionalHP 53131A/132A Performance Test Record Page 1 HP 53131A/132A Performance Test Record Page 1HP 53131A/132A Performance Test Record Page 2 HP 53131A/132A Performance Test Record Page 2HP 53131A/132A Performance Test Record Page 3 HP 53131A/132A Performance Test Record Page 3HP 53131A/132A Performance Test Record Page 4 HP 53131A/132A Performance Test Record Page 4HP 53131A/132A Performance Test Record Page 5 HP 53131A/132A Performance Test Record Page 5HP 53131A/132A Performance Test Record Page 6 HP 53131A/132A Performance Test Record Page 6HP 53181A Performance Tests HP 53181A Complete Performance TestsHP 53181A Complete Performance Tests Test 1 Trigger Level HP 53181A Only Test 1 Trigger Level HP 53181A OnlyHP 3325B HP 53181ASynthesizer Counter 11. Trigger Level Test SetupThis is the trigger level it should be 0.0 V ±15 mV Test 2 Channel 1 Frequency Sensitivity HP 53181A Only Test 2 Channel 1 Frequency Sensitivity HP 53181A OnlyKHz to 100 MHz Sensitivity for HP 53181A 12. Channels 1 Frequency Sensitivity Test SetupTest 2 Channel 1 Frequency Sensitivity HP 53181A Only MHz to 225 MHz Sensitivity for HP 53181A MHz to 200 MHz Sensitivity for HP 53181ATest 3 Channel 1 Frequency Accuracy HP 53181A Only Test 3 Channel 1 Frequency Accuracy HP 53181A OnlySweep Oscillator HP 8663A Synthesized Signal GeneratorHP 53181A Counter Test 3 Channel 1 Frequency Accuracy HP 53181A Only Equipment 14. Option 015/030/050 Channel 2 Frequency Test MHz to 1.5 GHz Sensitivity for Option 015 OnlyMHz to 2.7 GHz Sensitivity for Option 030 Only MHz to 5.0 GHz Sensitivity for Option 050 only MHz to 12.4 GHz Sensitivity for Option 124 only MHz to 1.5 GHz Accuracy Test for Option Connect equipment as shown in Figure Minimum to Maximum Accuracy Test for Options 030/050/124 Verify the HP 53181A Counter reads 100 MHz ±70 mHz Performance Tests Test 6 Peak Volts, Channel 1 HP 53181A Only Test 6 Peak Volts, Channel 1 HP 53181A Only16. Peak Volts Test Setup HP 53181A HP-IB Verification Program Optional HP 53181A HP-IB Verification Program OptionalHP 53181A Performance Test Record Page 1 HP 53181A Performance Test Record Page 1HP 53181A Performance Test Record Page 2 HP 53181A Performance Test Record Page 2HP 53181A Performance Test Record Page 3 HP 53181A Performance Test Record Page 3HP 53181A Performance Test Record Page 4 HP 53181A Performance Test Record Page 4Service Service To Provide Repair Information Returning the Instrument to Hewlett-Packard for ServiceReturning the Instrument to Hewlett-Packard for Service To Pack in the Original Packaging Materials To Pack in the Commercially Available Materials About the HP 53131A/132A Calibration Menu About the HP 53131A/132A Calibration MenuOverview of the HP 53131A/132A Calibration Menu CAL Unsecure HP 53131A/132A Calibration Menu Tree Calibration proceduresTo Initiate the Calibration Routines To View the Calibration Menu and Security StatusTo Unsecure for Calibration Information about Calti QUIK? and Calti FINE? About the HP 53131A/132A Calibration Menu HP 59992A J06 Time Interval Calibrator HP 8130A HP 53131A/132A Pulse Generator CounterTo Change to a New Security Code To Secure Against CalibrationTo Get Help with the Calibration Menu To View the Calibration CountHP 53131A/132A Calibration Procedures HP 53131A/132A Calibration ProceduresFirst Determine the Counter Firmware Revision To Calibrate the Offset for Channels 1BAD Code HP 53131A/132A Calibration Procedures Digital Multimeter DC Power Supply HP 53131A/132A CounterTo Calibrate the Gain for Channels 1 Press Enter key again To Calibrate Time IntervalHP 8130A HP53131A/132A Pulse Generator Counter CAL TI QUIK? CalibrationHP 8130A Pulse Generator CAL TI FINE? CalibrationTI CAL Pass House Standard MHz Reference HP 53131A/132A Counter To Calibrate the Standard TimebaseMedium, High, or Ultra-High To Calibrate the High Stability Timebase OptionAbout the HP 53181A Calibration Menu About the HP 53181A Calibration MenuOverview of the HP 53181A Calibration Menu About the HP 53181A Calibration Menu HP 53181A Calibration Menu Tree Key to access this menuTo View the Calibration Menu and Security Status To Initiate the Calibration Routines To Change to New the Security Code To Calibrate the Offset for Channel HP 53181A Calibration ProceduresHP 53181A Calibration Procedures BAD Code To Calibrate the Gain for Channel HP 53181A DC Power Supply Counter Digital MultimeterHouse Standard MHz Reference HP 53181A Counter To Calibrate the High Stability Timebase Option Pre-Troubleshooting Information Pre-Troubleshooting InformationSafety Considerations Interruption is Prohibited Recommended Test EquipmentRepair Considerations Disassembly and Reassembly SpecificsElectrostatic Discharge Surface Mount RepairProduct Performance Checks After Service ConsiderationsProduct Safety Checks HP 53131A Assembly Identification Assembly Identification and Location53132-60014 53131-69014 Assembly Name HP Part No Exchange Part A1 Motherboard Series 8A. Instrument Top Internal View 8B. Instrument Top Internal View Option Troubleshooting the Counter Power Supply CheckMotherboard Assembly Overview of the Self-Test Routines A4 AC Power Supply VoltagesWire Color Voltage Tolerance Test To Run the Test-All Self Test Diagnosing the Faulty Assembly by UsingSelf Tests Test Loop to OFF To Run the Individual Self Tests11B. HP 53181A Self Test Troubleshooting Setup Failure Indicator Assembly at Fault TroubleshootingReplacing Assemblies Replacing Assemblies Tools Required Tools RequiredDo This First Rear Bezel Removal To Remove the CoverBottom View for Cover Removal To Remove the Front Bezel To Remove the Front BezelFront Bezel Removal To Remove the Front Bezel To Remove A1 Motherboard Assembly To Remove A1 Motherboard AssemblyA1 Motherboard Assembly Removal To Remove the A2 Display Board, Keypads, and Window To Remove the A2 Display Board, Keypads, and WindowTo Remove the A2 Display Board, Keypads, and Window 015/030/050/124 To Remove A3 1.5/3.0/5.0/12.4 GHz Channel Assembly OptionReplacing Assemblies Optional Channel Input Assembly Removal To Remove A4 AC Power Supply Assembly To Remove A4 AC Power Supply Assembly10. Location of the Rubber Foot To Remove A5 DC Power Input Assembly Option To Remove A5 DC Power Input Assembly Option11. DC Power Input Assembly Option 002 Removal Remove the cover Remove A6 High Stability Timebase Assembly To Remove the Rear Terminals Option To Remove the Rear Terminals Option13. Rear Terminals Option 060 Removal To Remove the Rear Terminals Option To Remove the Rear Terminals Option Retrofitting Options Retrofitting Options Retrofitting Options Option 002 DC Power Input Assembly Parts To Retrofit A5 DC Power Input Assembly OptionPreliminary Procedure To Retrofit A5 DC Power Input Assembly Option1A. DC Power Input Assembly Retrofitting Retrofitting Procedure1B. DC Power Input Assembly Retrofitting To Retrofit A5 DC Power Input Assembly Option 1C. DC Power Input Assembly Retrofitting 1D. DC Power Input Assembly Retrofitting Dc-supply Option 012 Ultra-High Stability Timebase Assembly Parts Option 001 Medium Stability Timebase Assembly PartsOption 010 High Stability Timebase Assembly Parts Turn off the Counter and remove power cord Remove the cover High Stability Timebase Retrofitting Option 001, 010 To Retrofit the Rear Terminals Option To Retrofit the Rear Terminals OptionOption 060 Rear Terminals Parts 3A. Plug Removal for Rear Terminals Retrofitting 3B. Rear Terminals Retrofitting To the rear of the chassis To Retrofit the Rear Terminals Option Replaceable Parts Exchange Assemblies Exchange AssembliesReference Designations Reference DesignationsReplaceable Parts Parts Identification How To Order a PartHow To Order a Part For mail correspondence, use the address below Contacting Hewlett-PackardBy telephone Reference Designation Cabinet Parts and HardwareCabinet Parts and Hardware Code Mfr Mfr Part NumberOption MP2 MP3 Mfr Code Manufacturer Name Address Zip Code Manufacturer’s Code ListBackdating Instrument this Guide Directly Applies To Manual ChangesOlder Instruments Backdating Hardware Change 2 New A3 5/12.4 GHz Prescaler AssemblyChange 1 Below Series Prefix 3711A Backdating HardwareChange 3 New Tcxo boards 10. A4 AC Power Supply -9V Output Check Option 010/012 OnlyChange 4 Series 3710A to 3705A Change 6 Below Series Prefix 3416A Change 5 Series 3705A to 3546A, 3548AChange 8 Below Serial Prefix 3404A Change 7 Below Serial Number 3349A01906About the HP 53131A Calibration Menu Backdating FirmwareChange 1 Firmware Revisions 3317, 3335, 3402 Backdating FirmwareAbout the HP 53131A Calibration Menu HP 53131A Utility/Calibration Menu TreeHP 53131A Calibration Procedures HP 53131A Calibration ProceduresDC Power Supply HP 53131A Counter To Calibrate Time Interval HP 8130A HP53131APulse Generator Counter EquipmentHouse Standard MHz Reference HP 53131A Counter Medium or High Backdating Specifications Backdating SpecificationsChange 1 53131A, Serial Prefix 3710A and Below Change 2 53132A, Serial Prefix 3546A and BelowChange 3 53181A, Serial Prefix 3711A02841 and Below Backdating Specifications HP 53131A/132A Specifications Instrument Inputs Channel 1 & 2 Input Specifications1Channel 1 & 2 Input Characteristics1 Channel 3 Input Characteristics Channel 3 Input Specifications4External Arm Input Specifications External Arm Input CharacteristicsTime Base Time Base Output SpecificationsTime Base Internal Time Base StabilityFrequency, Period Measurement SpecificationsMeasurement Specifications Automatic or External Arming HP 53131A-Worst Case RMS ResolutionHP 53132A-Worst Case RMS Resolution × Frequency Frequency Measurement ExampleLSD Time IntervalRise/Fall Time PhasePulse Width Peak Volts Duty CycleTotalize Trigger Level Timing Error Definitions of Systematic Uncertainty Terms Trigger ErrorMeasurement Definitions Measurement DefinitionsAdditional Considerations for Digits and Time Arming Modes Typical Versus Worst Case SpecificationsTime Interval Measurement LSD Definition Measurement Arming and Processing Measurement Arming and ProcessingGate Time Measurement Throughput Measurement ArmingMeasurement Limits Measurement StatisticsGeneral Information General InformationHP 53181A Specifications Channel 1 Input Characteristics Channel 1 Input SpecificationsChannel 2 Input Characteristics Channel 2 Input Specifications3Time Base Where N = 1 for Ch1 Frequency 1 MHz Worst Case RMS Resolution Frequency Measurement Example Ratio 1 ⁄ Ch2 Freq × Gate Time Ratio 2 ⁄ Typical Versus Worst Case Specifications Available Statistics Power is removed from the counter General Information Index Index-2 Assembly-Level Service Guide Assembly-Level Service Guide Index-3 Index-4 Assembly-Level Service Guide Contacting Hewlett-Packard Page From front matter Manual Part Number