HP 181A manual External Timebase Tests

Page 34

1

Chapter 1 Performance Tests

HP 53131A/132A/181A Operational Verification

3Set the HP 8663A to output a 200 MHz, 100 mV sine wave signal to Channel 1.

4On the Counter, Press Run key.

5Send a single pulse by pressing the MAN key on the HP 8130A, and observe that with each press the Gate annunciator flashes.

The Counter should display approximately 1000.

6For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A Performance Test Record on page 1-41, Test 7.

7Disconnect the test setup.

External Timebase Tests

These tests verify the External Timebase 1, 5, 10 MHz specifications of the HP 53131A and HP 53181A Counters by checking that TIMEBASE AUTO allows the Counter to be synchronized to the external reference.

Note that the HP 53132A only operates with the 10 MHz external timebase; therefore, perform the procedure in the section titled “10 MHz External Timebase Input for the HP 53132A” on page 1-15 to verify that TIMEBASE AUTO allows the Counter to be synchronized to the external reference.

Equipment

HP 3325B Synthesizers (2)

HP 10100C 50W Feedthrough

HP 10503 BNC Cables (2)

1-12

Assembly-Level Service Guide

Image 34
Contents HP 53131A/132A/181A 225 MHz Counter Page Assembly-Level Service Guide Safety Considerations contd Contents Test 2 Trigger Level HP 53131A/132A Only Test 5 Option 030/050/124 Channel 3 Frequency Test 2 Channel 1 Frequency Sensitivity HP 53181A Service About the HP 53181A Calibration Menu Replacing Assemblies Retrofitting Options Index Contents Xii Assembly-Level Service Guide Repair Strategy How to Use This GuideInstrument Identification Assembly-Level Service Guide Organization Instruments Covered by this GuideHow to Order Guides Description of the 225 MHz Counters Assembly-Level Service Guide Xvii Options HardwareAccessories Available Accessories Supplied and AvailableAccessories Supplied SupportPreface Performance Tests Operational Verification IntroductionIntroduction Recommended Calibration Cycle Complete Performance TestsOptional HP-IB Verification Test RecordEquipment Required Equipment RequiredRecommended Test Equipment HP 53131A/132A/181A Operational Verification Power-On Self TestsHP 53131A/132A/181A Operational Verification HP 53131A/132A Counter1B. Operational Verification Setup for HP 53181A HP 53131A/132A/181A Operational Verification HP 53181ACounter HP 3458A Run Self TestTermination Check Digital MultimeterPress Channel 1 50Ω/1MΩ key External Arm Test HP 53131A/132A Only Counter SetupEquipment HP 8130A Pulse Generator ProcedureSynthesized Signal Generator HP 8663AExternal Timebase Tests HP 3325B Synthesizer #1 MHz External Timebase Input HP 53131A and HP 53181A OnlyHP 3325B Synthesizer #2 Counter MHz External Timebase Input HP 53131A and HP 53181A Only MHz External Timebase Input for the HP 53132A HP 53131A/132A Complete Performance Tests HP 53131A/132A Complete Performance TestsHP 53131A/132A Performance Tests Equipment Counter SetupTest 1 Time Interval HP 53131A/132A Only Test 1 Time Interval HP 53131A/132A OnlyProcedure HP 8130A HP53131A/132Pulse Generator CounterTest 2 Trigger Level HP 53131A/132A Only Test 2 Trigger Level HP 53131A/132A OnlyTrigger Level Test Setup HP 3325B HP 53131A/132A Synthesizer CounterThis is the trigger level it should be 0.0 V ±15 mV Equipment HP 53131A/132A Ref. Counter HP 8663A Synthesized Signal GeneratorKHz to 100 MHz Sensitivity for HP 53131A/132A HP 8663A SynthesizedPerformance Tests MHz to 200 MHz Sensitivity for HP 53131A/132A MHz to 225 MHz Sensitivity for HP 53131A/132A Equipment Channels 1 and 2 Frequency Accuracy Test Setup HP 53131A/132A CounterPerformance Tests Equipment HP 53131A/132A Counter HP 8663A Synthesized Signal GeneratorSweep Oscillator To 3.0 GHz Sensitivity for Option 030 only Sweep the frequencies from 200 MHz to 5.0 GHz Sweep the frequencies from 200 MHz to 12.4 GHz Gate & ExtArm, Time Time 1.000 s On the Counter, press Run key Test 7 Peak Volts, Channels 1 and 2 HP 53131A/132A Only Test 7 Peak Volts, Channels 1 and 2 HP 53131A/132A OnlyVolt Peak HP 3325B HP 53131A/132ASynthesizer CounterVolt Peak HP 53131A/132A HP-IB Verification Program Optional HP 53131A/132A HP-IB Verification Program OptionalHP 53131A/132A Performance Test Record Page 1 HP 53131A/132A Performance Test Record Page 1HP 53131A/132A Performance Test Record Page 2 HP 53131A/132A Performance Test Record Page 2HP 53131A/132A Performance Test Record Page 3 HP 53131A/132A Performance Test Record Page 3HP 53131A/132A Performance Test Record Page 4 HP 53131A/132A Performance Test Record Page 4HP 53131A/132A Performance Test Record Page 5 HP 53131A/132A Performance Test Record Page 5HP 53131A/132A Performance Test Record Page 6 HP 53131A/132A Performance Test Record Page 6HP 53181A Complete Performance Tests HP 53181A Complete Performance TestsHP 53181A Performance Tests Test 1 Trigger Level HP 53181A Only Test 1 Trigger Level HP 53181A Only11. Trigger Level Test Setup HP 3325B HP 53181ASynthesizer CounterThis is the trigger level it should be 0.0 V ±15 mV Test 2 Channel 1 Frequency Sensitivity HP 53181A Only Test 2 Channel 1 Frequency Sensitivity HP 53181A Only12. Channels 1 Frequency Sensitivity Test Setup KHz to 100 MHz Sensitivity for HP 53181ATest 2 Channel 1 Frequency Sensitivity HP 53181A Only MHz to 200 MHz Sensitivity for HP 53181A MHz to 225 MHz Sensitivity for HP 53181ATest 3 Channel 1 Frequency Accuracy HP 53181A Only Test 3 Channel 1 Frequency Accuracy HP 53181A OnlyHP 53181A Counter HP 8663A Synthesized Signal GeneratorSweep Oscillator Test 3 Channel 1 Frequency Accuracy HP 53181A Only Equipment MHz to 1.5 GHz Sensitivity for Option 015 Only 14. Option 015/030/050 Channel 2 Frequency TestMHz to 2.7 GHz Sensitivity for Option 030 Only MHz to 5.0 GHz Sensitivity for Option 050 only MHz to 12.4 GHz Sensitivity for Option 124 only MHz to 1.5 GHz Accuracy Test for Option Connect equipment as shown in Figure Minimum to Maximum Accuracy Test for Options 030/050/124 Verify the HP 53181A Counter reads 100 MHz ±70 mHz Performance Tests Test 6 Peak Volts, Channel 1 HP 53181A Only Test 6 Peak Volts, Channel 1 HP 53181A Only16. Peak Volts Test Setup HP 53181A HP-IB Verification Program Optional HP 53181A HP-IB Verification Program OptionalHP 53181A Performance Test Record Page 1 HP 53181A Performance Test Record Page 1HP 53181A Performance Test Record Page 2 HP 53181A Performance Test Record Page 2HP 53181A Performance Test Record Page 3 HP 53181A Performance Test Record Page 3HP 53181A Performance Test Record Page 4 HP 53181A Performance Test Record Page 4Service Service Returning the Instrument to Hewlett-Packard for Service Returning the Instrument to Hewlett-Packard for ServiceTo Provide Repair Information To Pack in the Original Packaging Materials To Pack in the Commercially Available Materials Overview of the HP 53131A/132A Calibration Menu About the HP 53131A/132A Calibration MenuAbout the HP 53131A/132A Calibration Menu CAL Unsecure Calibration procedures HP 53131A/132A Calibration Menu TreeTo Unsecure for Calibration To View the Calibration Menu and Security StatusTo Initiate the Calibration Routines Information about Calti QUIK? and Calti FINE? About the HP 53131A/132A Calibration Menu HP 8130A HP 53131A/132A Pulse Generator Counter HP 59992A J06 Time Interval CalibratorTo Secure Against Calibration To Change to a New Security CodeTo View the Calibration Count To Get Help with the Calibration MenuFirst Determine the Counter Firmware Revision HP 53131A/132A Calibration ProceduresHP 53131A/132A Calibration Procedures To Calibrate the Offset for Channels 1BAD Code HP 53131A/132A Calibration Procedures To Calibrate the Gain for Channels 1 DC Power Supply HP 53131A/132A CounterDigital Multimeter To Calibrate Time Interval Press Enter key againCAL TI QUIK? Calibration HP 8130A HP53131A/132A Pulse Generator CounterCAL TI FINE? Calibration HP 8130A Pulse GeneratorTI CAL Pass To Calibrate the Standard Timebase House Standard MHz Reference HP 53131A/132A CounterTo Calibrate the High Stability Timebase Option Medium, High, or Ultra-HighOverview of the HP 53181A Calibration Menu About the HP 53181A Calibration MenuAbout the HP 53181A Calibration Menu About the HP 53181A Calibration Menu Key to access this menu HP 53181A Calibration Menu TreeTo View the Calibration Menu and Security Status To Initiate the Calibration Routines To Change to New the Security Code HP 53181A Calibration Procedures HP 53181A Calibration ProceduresTo Calibrate the Offset for Channel BAD Code HP 53181A DC Power Supply Counter Digital Multimeter To Calibrate the Gain for ChannelHouse Standard MHz Reference HP 53181A Counter To Calibrate the High Stability Timebase Option Safety Considerations Pre-Troubleshooting InformationPre-Troubleshooting Information Recommended Test Equipment Interruption is ProhibitedElectrostatic Discharge Disassembly and Reassembly SpecificsRepair Considerations Surface Mount RepairProduct Safety Checks After Service ConsiderationsProduct Performance Checks Assembly Identification and Location HP 53131A Assembly Identification53132-60014 53131-69014 Assembly Name HP Part No Exchange Part A1 Motherboard Series 8A. Instrument Top Internal View 8B. Instrument Top Internal View Option Power Supply Check Troubleshooting the CounterMotherboard Assembly Wire Color Voltage Tolerance A4 AC Power Supply VoltagesOverview of the Self-Test Routines Test Self Tests Diagnosing the Faulty Assembly by UsingTo Run the Test-All Self Test To Run the Individual Self Tests Test Loop to OFF11B. HP 53181A Self Test Troubleshooting Setup Troubleshooting Failure Indicator Assembly at FaultReplacing Assemblies Replacing Assemblies Do This First Tools RequiredTools Required To Remove the Cover Rear Bezel RemovalBottom View for Cover Removal To Remove the Front Bezel To Remove the Front BezelFront Bezel Removal To Remove the Front Bezel To Remove A1 Motherboard Assembly To Remove A1 Motherboard AssemblyA1 Motherboard Assembly Removal To Remove the A2 Display Board, Keypads, and Window To Remove the A2 Display Board, Keypads, and WindowTo Remove the A2 Display Board, Keypads, and Window Replacing Assemblies To Remove A3 1.5/3.0/5.0/12.4 GHz Channel Assembly Option015/030/050/124 Optional Channel Input Assembly Removal To Remove A4 AC Power Supply Assembly To Remove A4 AC Power Supply Assembly10. Location of the Rubber Foot To Remove A5 DC Power Input Assembly Option To Remove A5 DC Power Input Assembly Option11. DC Power Input Assembly Option 002 Removal Remove the cover Remove A6 High Stability Timebase Assembly To Remove the Rear Terminals Option To Remove the Rear Terminals Option13. Rear Terminals Option 060 Removal To Remove the Rear Terminals Option To Remove the Rear Terminals Option Retrofitting Options Retrofitting Options Retrofitting Options Preliminary Procedure To Retrofit A5 DC Power Input Assembly OptionOption 002 DC Power Input Assembly Parts To Retrofit A5 DC Power Input Assembly OptionRetrofitting Procedure 1A. DC Power Input Assembly Retrofitting1B. DC Power Input Assembly Retrofitting To Retrofit A5 DC Power Input Assembly Option 1C. DC Power Input Assembly Retrofitting 1D. DC Power Input Assembly Retrofitting Dc-supply Option 010 High Stability Timebase Assembly Parts Option 001 Medium Stability Timebase Assembly PartsOption 012 Ultra-High Stability Timebase Assembly Parts Turn off the Counter and remove power cord Remove the cover High Stability Timebase Retrofitting Option 001, 010 Option 060 Rear Terminals Parts To Retrofit the Rear Terminals OptionTo Retrofit the Rear Terminals Option 3A. Plug Removal for Rear Terminals Retrofitting 3B. Rear Terminals Retrofitting To the rear of the chassis To Retrofit the Rear Terminals Option Replaceable Parts Exchange Assemblies Exchange AssembliesReplaceable Parts Reference DesignationsReference Designations How To Order a Part How To Order a PartParts Identification By telephone Contacting Hewlett-PackardFor mail correspondence, use the address below Cabinet Parts and Hardware Cabinet Parts and HardwareReference Designation Mfr Mfr Part Number CodeOption MP2 MP3 Manufacturer’s Code List Mfr Code Manufacturer Name Address Zip CodeBackdating Older Instruments Manual ChangesInstrument this Guide Directly Applies To Change 1 Below Series Prefix 3711A Change 2 New A3 5/12.4 GHz Prescaler AssemblyBackdating Hardware Backdating Hardware10. A4 AC Power Supply -9V Output Check Option 010/012 Only Change 3 New Tcxo boardsChange 4 Series 3710A to 3705A Change 5 Series 3705A to 3546A, 3548A Change 6 Below Series Prefix 3416AChange 7 Below Serial Number 3349A01906 Change 8 Below Serial Prefix 3404AChange 1 Firmware Revisions 3317, 3335, 3402 Backdating FirmwareAbout the HP 53131A Calibration Menu Backdating FirmwareHP 53131A Utility/Calibration Menu Tree About the HP 53131A Calibration MenuHP 53131A Calibration Procedures HP 53131A Calibration ProceduresDC Power Supply HP 53131A Counter To Calibrate Time Interval Equipment HP 8130A HP53131APulse Generator CounterHouse Standard MHz Reference HP 53131A Counter Medium or High Change 1 53131A, Serial Prefix 3710A and Below Backdating SpecificationsBackdating Specifications Change 2 53132A, Serial Prefix 3546A and BelowChange 3 53181A, Serial Prefix 3711A02841 and Below Backdating Specifications HP 53131A/132A Specifications Channel 1 & 2 Input Characteristics1 Channel 1 & 2 Input Specifications1Instrument Inputs External Arm Input Specifications Channel 3 Input Specifications4Channel 3 Input Characteristics External Arm Input CharacteristicsTime Base Time Base Output SpecificationsTime Base Internal Time Base StabilityMeasurement Specifications Measurement SpecificationsFrequency, Period HP 53131A-Worst Case RMS Resolution Automatic or External ArmingHP 53132A-Worst Case RMS Resolution Frequency Measurement Example × FrequencyTime Interval LSDPulse Width PhaseRise/Fall Time Totalize Duty CyclePeak Volts Measurement Definitions Definitions of Systematic Uncertainty Terms Trigger ErrorTrigger Level Timing Error Measurement DefinitionsTime Interval Measurement LSD Definition Typical Versus Worst Case SpecificationsAdditional Considerations for Digits and Time Arming Modes Gate Time Measurement Throughput Measurement Arming and ProcessingMeasurement Arming and Processing Measurement ArmingMeasurement Statistics Measurement LimitsGeneral Information General InformationHP 53181A Specifications Channel 1 Input Specifications Channel 1 Input CharacteristicsChannel 2 Input Specifications3 Channel 2 Input CharacteristicsTime Base Where N = 1 for Ch1 Frequency 1 MHz Worst Case RMS Resolution Frequency Measurement Example Ratio 1 ⁄ Ch2 Freq × Gate Time Ratio 2 ⁄ Typical Versus Worst Case Specifications Available Statistics Power is removed from the counter General Information Index Index-2 Assembly-Level Service Guide Assembly-Level Service Guide Index-3 Index-4 Assembly-Level Service Guide Contacting Hewlett-Packard Page From front matter Manual Part Number