Section 3 - Operation

3.1 Power Requirements

Caution! Be sure that the power supply is the same voltage as specified on the nameplate

With the chiller plugged in (before pressing the POWER ON button), the display responds by showing standby (....). If there is no response, check that the circuit breaker on the rear of the chiller is in the

ON position. An extension cord is not recommended.

(For models with

 

 

 

LED

°C OR °F

 

SETTING

optional heater only)

 

 

 

DISPLAY

DISPLAY

ADJUSTMENT KNOB

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Front Face Plate

PRESSURE

 

 

GAUGE,

SET/MENU

POWER

 

(On Positive Displacement

BUTTON

BUTTONS

 

 

Pump models only)

 

 

 

 

 

 

 

 

3.2Temperature Setting and Adjustments

After setting up and filling the chiller:

1.Press the POWER ON button. The LED display indicates (8888), the power up self test. When the chiller is turned on, there is a 60 second time-out period. During this time the controller is performing a self-diagnostic and a balancing of pressure across the compressor inlet and outlet. (a balanced system reduces the AC current required to start up the compressor and will result in an extended compressor life.) The pump is active during the start delay period. When the time out period reaches zero, the compressor starts and the chiller is then fully operational.

Upon completion of the self test, the actual fluid temperature is displayed. Your desired operating temperature can now be set.

2.Press the SET/MENU button. The degree lights flash indicating the temperature can be set or changed. Turn the SETTING ADJUSTMENT knob to the desired setting. This setting is accepted after pressing the SET/MENU button or will be automatically accepted after a few seconds of keypad inactivity. The degree light stops flashing and the display shows the actual temperature.

3.The SET TEMPERATURE may be checked at any time by pressing and releasing the SET/MENU button. Allow sufficient time for the chiller to stabilize at the desired temperature.

3.3Selection of Celsius or Fahrenheit

To change the readout to °C or °F:

1.Press and hold the Set/Menu button until the display reads “UNITS”.

2.Press Set/Menu again then turn the Setting Adjustment knob and select °C or °F.

3.Press Set/Menu or the setting will be accepted after a few seconds.

9

Image 9
Contents Operators Manual Refrigerated Recirculating Chillers Table of Contents General Information Package Contents WarrantyGeneral Information UnpackingGeneral Specifications To All Chillers Chiller Specifications and Pump PerformanceModel 6205 6305 6505 Model 6705 6105 6155 Set Up Before proceeding, be sure all power is off LocationReservoir Fluids Setting Up an Open Bath System to the Chiller Setting Up a Closed System or Cooling Coil to the ChillerTemperature Setting and Adjustments Power RequirementsOperation Selection of Celsius or FahrenheitController Default Settings Setting the Readout High LimitController Display Messages Draining the Unit6705 Diagram of Models6155 Positive Displacement and Turbine High Pressure Pump Motor MaintenanceStandard Magnetic Drive Centrifugal Pump Motor Fluid Filter Included With High Pressure Pump