Page 10

3.4 Setting the Readout High Limit

This feature limits the readout setability so the operating temperature setting cannot easily be set above the High Limit you select. It is not an over temperature feature which shuts off the chiller.

To set the readout High Limit:

1.Press and hold the Set/Menu button until the display reads “UNITS”.

2.Turn the Setting Adjustment knob until the display reads “HI-L”.

3.Press the Set/Menu button, enter the desired value using the Setting Adjustment knob.

4.Press Set/Menu and the display will again read "HI-L". The Setting Adjustment knob can be turned to access another menu option or you can wait until the entered value has been accepted. When the value is accepted, the degree light will stop flashing and the fluid temperature will be displayed.

3.5Controller Display Messages

Display

Message

 

--Normal Screens--

 

 

....

Standby mode. Unit plugged in. Power switch OFF, rear circuit breaker on

 

 

8888

Power up self test displayed when power button initially pressed on

 

 

 

--Menu Screen--

 

 

Unit

Change to °F or °C

 

 

Unit/Hi-L

Limits operating temperature setability

 

 

 

--Error Screens--

 

 

E-tF

Triac (alternistor) failure

 

 

DEF

Displayed when chiller is reset to factory default settings

 

 

 

--Optional Equipment (Flow or Float Switch)--

 

 

E-nF

No Flow

E-FL

Low Level Fluid

 

 

3.6 Draining the Unit

Press the Power Off button and unplug the chiller. Remove drain plug and reservoir cap and allow unit to drain. Syphon or use a small manual pump to remove all fluid from the reservoir.

Run the unit intermittently (5 to 10 seconds each time) to pump the water from the system. Stop pumping when the fluid is reduced to a trickle. Running for a longer period will cause damage to the positive displacement pump. Open the reservoir and use suction or a hand pump to remove any remaining water.

Prior to storage or shipping: Water MUST be removed from ALL parts of the chiller system, or an ANTIFREEZE must be added to insure that internal damage does not occur.

3.7 Controller Default Settings

To return to the factory default values:

1.Press and hold the Power On button, then turn off at the circuit breaker.

2.Still holding Power On, hold the Set/Menu button while turning the unit on by the circuit breaker.

3.The controller will display “DEF” and go to the standby mode.

4.Your set temperature and other settings will have to be reset.

10

Image 10
Contents Operators Manual Refrigerated Recirculating Chillers Table of Contents General Information General Information WarrantyPackage Contents UnpackingChiller Specifications and Pump Performance General Specifications To All ChillersModel 6205 6305 6505 Model 6705 6105 6155 Before proceeding, be sure all power is off Location Set UpReservoir Fluids Setting Up a Closed System or Cooling Coil to the Chiller Setting Up an Open Bath System to the ChillerOperation Power RequirementsTemperature Setting and Adjustments Selection of Celsius or FahrenheitController Display Messages Setting the Readout High LimitController Default Settings Draining the UnitDiagram of Models 67056155 Standard Magnetic Drive Centrifugal Pump Motor MaintenancePositive Displacement and Turbine High Pressure Pump Motor Fluid Filter Included With High Pressure Pump