MΩ (Resistance)

 

 

 

 

Range

Resolution

Accuracy

Overload Protection

9.999MΩ

1KΩ

±(5% + 10d)

600V

99.99MΩ

10KΩ

 

 

 

 

 

 

Capacitance

 

 

 

Range

Resolution

Accuracy

Overload Protection

10.000μF

1nF

 

 

100.00μF

10nF

±(1.5% + 5d)

600V

1000.0μF

100nF

 

 

 

7000μF

1μF

±(2.5% + 15d)

 

 

 

 

 

 

 

Diode (Continuity <40mV)

 

 

Range

Resolution

Accuracy

Overload Protection

2.000V

1mV

±(2% + 1d)

600V

 

 

 

 

 

 

 

Temperature (K-Type thermocouple)

 

 

Range

Resolution

Accuracy

Overload Protection

-50oC to 900oC

0.1oC

±(1% + 1oC)

30VAC or 60VDC

-58oF to 1000oF

0.1oF

±(1% + 2oF)

 

 

 

1Φ/3Φ TRUE Power (PF > 0.5 or < 60o) (50/60Hz)

 

Range

Resolution

Accuracy

Overload Protection

60.00KW (<100A)

10W

±(5% + 20d)

600VAC/

600.0KW (>100A)

100W

±(5% + 20d)

1000AAC

 

 

 

 

 

 

 

1Φ/3Φ Horse Power (HP) (PF > 0.5 or

< 60o) (50/60Hz)

 

Range

Resolution

Accuracy

Overload Protection

80.00HP (<100A)

0.01 HP

±(5% + 20d)

600VAC/1000AAC

800.0 HP (>100A)

0.1 HP

±(5% + 20d)

 

 

 

 

 

 

1Φ/3Φ Reactive Power (KVAR) (PF > 0.5 or < 60o) (50/60Hz)

Range

Resolution

Accuracy

Overload Protection

60.00KVAR (<100A)

10VAR

±(5% + 20d)

600VAC/

600.0KVAR (>100A)

100VAR

±(5% + 20d)

1000AAC

 

 

 

 

19

380976

V3.1

6/08

Image 19
Contents 1Φ/3Φ 1000 Amp True RMS Power Clamp-On Model Introduction Table of ContentsSafety Information WarrantyWhen servicing, use only specified replacement parts Meter Description AC + DC Voltage Measurements MeasurementsAC Current Measurements 380976 V3.1 380976 V3.1 KVA 3φ 3W KW+PF Power Factor, KW+KVAR, KVA+θPhase Angle or A+V380976 V3.1 380976 V3.1 380976 V3.1 380976 V3.1 Resistance and Audible Continuity Measurements KVA 3φ 4WAC and DC µA Measurements Temperature MeasurementsCapacitance Measurements Diode TestsHold MX/MN Key Meter Control KeysRange Key Auto Sleep Disable Automatic Sleep Mode and Battery ReplacementBattery Replacement Specifications AC Voltage 50Hz to 400Hz True RMS Measurement SpecificationsAC Current 50Hz to 400Hz True RMS ΜA True RMS AC+DC1Φ/3Φ Reactive Power Kvar PF 0.5 or 60o 50/60Hz Temperature K-Type thermocouple1Φ/3Φ True Power PF 0.5 or 60o 50/60Hz 1Φ/3Φ Horse Power HP PF 0.5 or 60o 50/60HzFrequency Calibration and Repair Services1Φ/3Φ Apparent Power KVA Phase Angle 50/60Hz