Page 35
Model 738AR

TM 11-6625-1614-15

Model 410C

Section V

Paragraphs

5-1to 5-11

SECTION V

MAINTENANCE

5 - 1 . I N T R O D U C T I O N .

 

5 - 9 . M e c h a n i c a l M E T E R Z E R O .

5-2.

This section contains

maintenance procedures

 

 

 

 

a. Turn instrument on.

Allow at least a 20

for

the Model 410C Electronic Voltmeter.

minute warm-up period.

 

 

5-3.

TEST EQUIPMENT

REQUIRED.

 

 

 

 

 

 

5-4. The test equipment required to maintain and adjust the Model 410C is listed in Table 5-1. Equi- pment having similar characteristics may be substi- tuted for items listed.

5 - 5 . PERFORMANCE CHECKS .

5 - 6 . The performance checks presented in this section are front panel operations designed to com- pare the Model 410C with it’s published specifications. These operations may be incorporated in periodic maintenance, post repair and incoming quality control checks. These operations should be conducted before any attempt is made at instrument calibration or adjustment. During performance checks, periodically vary the line voltage to the Model 410C, ± 10% on either 115v or 230 v operation. A 1/2 hour warm-up period should be allowed before these tests are conducted.

5-7. ALTERNATE VOLTAGE SOURCE.

5-6. Should it be necessary to use the

Voltmeter Calibrator to conduct these Performance Checks, the arrangement described in Figure 5-1will provide the necessary voltage values required. How- ever; the Model 738BR Voltmeter Calibrator is the preferred instrument for these operations.

b. Turn voltmeter off, and allow 30 seconds for all capacitors to discharge.

c.Rotate mechanical zero-adjustment screw on front panel clockwise until pointer reaches zero, moving up scale.

d.If for some reason the pointer should over- shoot zero, repeat step c until desired results are obtained.

e. When pointer has been positioned at zero, rotate zero-adjust screw slightly counterclock- wise to free it. If meter pointer moves to the left during this action, repeat steps c and e.

5-10. DC VOLTMETER OPERATION.

5-11. ACCURACY CHECK (DCV).

a. Set the Model 410C FUNCTION SELECTOR to the +DCV position; RANGE switch to. 015 V .

Connect Model 410C DCV and COM cables to the

Voltmeter Calibrator Model 738BR) output terminals .

Figure 5-1. Alternate Voltage Source

01556-2

5-1

Image 35
Contents HEWLETT-PACKARD Electronic Voltmeter Model 410C Exist in this Equipment HEWLETT-PACKARD Electronic Voltmeter List of Tables Number Section Section General Information Page Model 410C Description Model 410C Figures 2-land Section Installation 01556-1 Section Operation Front and Rear Panel Controls 01556-2 Pulse Measurements Voltage Measurement AT High FRENegative Pulses PRFDC Voltage Measurements 01556-3 AC Voltage Measurements Section Model 410C 5 5 6 Graph Used in Calculation of Pulse Voltage Readings IIIResistance Measurements 01556-2 DC Nano-Amoere Current Measurements Section MODULATOR-DEMODULATOR Input NetworkFeedback Network Power SupplySection 01556-2 Section Section 01556-2 Section Page BNC DC Voltmeter Operation Accuracy Check DCV Alternate Voltage SourceInput Resistance Check DCA Input Resistance Check DCVModel 410C Section Tables 5-3and Amplifier Gain Check Amplifier OperationOhmmeter Operation AC Voltmeter Operation AC Probe Accuracy Check DC Amplifter Output Impedance CheckAC Voltmeter Frequency Response Check AC Voltmeter Accuracy CheckSET Level Chopper Frequency Adjust Adjustment and Calibration ProcedureCOM N C T I O NPage C e t h e y a r e w i t h i n Troubleshooting Procedure C Zero AdjustAC Full Scale Adjust .5 V Range D e l 4 1 0 C Servicing Etched Circuit Boards Troubleshooting Tree 01566-2Chopper Assembly Installation Paragraph Section Model 410C 11. Model 11036A AC Probe Exploded View Change 16.1 Page Section Power Supply Schematic Typical Amplifier Waveforms Page Page Page Harold K. Johnson Page Metric System and Equivalents N 0 1 6 2 8 8 0 0