HP 410C manual AC Voltmeter Accuracy Check, AC Voltmeter Frequency Response Check

Page 40

TM 11-6625-1614-15

Section

V

Model 410C

Parsgraphs 5-24 to 5-25

Figure

5-4,

Table 5-6

Figure 5-4. Low Frequency Response Test

AC VOLTMETER ACCURACY CHECK.

a. A

Voltmeter

Calibrator

Model 738BR)

will

be required

for this

operation.

b. Adjust voltmeter calibrator for 400 cps- rms output.

c. Set Model 410C FUNCTION SELECTOR to ACV; RANGE to 500 V.

d. Adjust the voltmeter calibrator to settings listed in Table 5-6.Model 410C should indicate readings within limits specified. If not, refer

5-25. AC VOLTMETER FREQUENCY RESPONSE CHECK.

a. A Frequency Response Test Set Model

739AR), a Test Oscillator Model 651A), an

RF Signal Generator

Model 608 C), a Power

Meter

Model 431

B), a Thermistor

Mount

Model 478A), a Probe - T - Connector

Model 11042A), a VHF Signal Generator

Model 612A) and a 10

KC Filter

Model K02-

411A)

will be required for this operation. Fig-

ure 5-3

and 5-4 describe the arrangement

to be

Table 5-6. AC Accuracy Test

5-6

C h a n g e 1

01556-2

Image 40
Contents HEWLETT-PACKARD Electronic Voltmeter Model 410C Exist in this Equipment HEWLETT-PACKARD Electronic Voltmeter List of Tables Number Section Section General Information Page Model 410C Description Model 410C Figures 2-land Section Installation 01556-1 Section Operation Front and Rear Panel Controls 01556-2 Voltage Measurement AT High FRE Pulse MeasurementsPRF Negative PulsesDC Voltage Measurements 01556-3 AC Voltage Measurements Section Model 410C 5 5 6 III Graph Used in Calculation of Pulse Voltage ReadingsResistance Measurements 01556-2 DC Nano-Amoere Current Measurements Section Input Network MODULATOR-DEMODULATORPower Supply Feedback NetworkSection 01556-2 Section Section 01556-2 Section Page BNC Alternate Voltage Source DC Voltmeter Operation Accuracy Check DCVInput Resistance Check DCV Input Resistance Check DCAModel 410C Section Tables 5-3and Ohmmeter Operation Amplifier OperationAmplifier Gain Check DC Amplifter Output Impedance Check AC Voltmeter Operation AC Probe Accuracy CheckAC Voltmeter Accuracy Check AC Voltmeter Frequency Response CheckSET Level Adjustment and Calibration Procedure Chopper Frequency AdjustN C T I O N COMPage C e t h e y a r e w i t h i n AC Full Scale Adjust .5 V Range C Zero AdjustTroubleshooting Procedure D e l 4 1 0 C Troubleshooting Tree 01566-2 Servicing Etched Circuit BoardsChopper Assembly Installation Paragraph Section Model 410C 11. Model 11036A AC Probe Exploded View Change 16.1 Page Section Power Supply Schematic Typical Amplifier Waveforms Page Page Page Harold K. Johnson Page Metric System and Equivalents N 0 1 6 2 8 8 0 0