Teledyne manual M201A Ammonia Analyzer, Principle of Operation

Page 7

Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1

1.0 M201A AMMONIA ANALYZER

This manual is to be used in conjunction with the M200A NOx Analyzer manual p/n 02246 supplied with this instrument.

The M201A Ammonia Analyzer consists of a modified M200A Nitrogen Oxides Analyzer and M501NH Ammonia Converter:

A functional block diagram of the system is shown in Figure 1.

The instrument comes in three pneumatic configurations:

The basic system is shown in Figure 3;

The M201A with Z/S valves is shown in Figure 4;

The M201A with External Calibrator Option (ECO) is shown in Figure 5.

Figure 2 can be used to connect the M201A system for installation. Please note that for a M201A with either a Zero/Span or ECO option, you must connect the umbilical cord, 02255, between the analyzer and the converter assembly.

1.1 Principle of Operation

The Teledyne API Model 201A Analyzer measures ammonia by converting NH3 to nitric oxide by the following reaction:

4NH3 + 5O2 → 4NO + 6H2O

The resulting NO is then measured by the chemiluminescent reaction of NO with ozone. Consult Section 4.1 in the M200A manual (provided with this instrument) for more details on NO measurement.

A molybdenum converter is continuously in the pneumatic pathway, providing the TN channel signal. The M501NH ammonia converter is periodically switched so both converters are in series, creating the TNx signal.

The ammonia concentration is computed by the difference between the TNx and TN channels, as shown in the following equation:

NH3 → TNx - TN

See the pneumatic diagram in Figure 1.

The actual formula for computation of ammonia concentration is more complicated than the above equation, as it includes converter efficiencies, plus sample temperature and pressure. The difficulty in ammonia measurement is that the molybdenum converter

6

Image 7
Contents Model 201A Ammonia Analyzer Safety Messages Table of Contents List of Figures List of Tables Page Principle of Operation M201A Ammonia AnalyzerNo → no Special Considerations for Ammonia MeasurementSample Filtration M201A Analyzer Specifications Zero/Span Calibration Basic Instrument Calibration ProcedureExit CALZero EntrCAL-CONV-SET Calz Zero/Span Calibration with Z/S Valves or ECO OptionTNX ConcCONC-TN Cals 2 TN/TNx Channel Calibration QualitySpan 3 NH3 Calibration Channel Calibration Quality M201A Maintenance Schedule Maintenance ScheduleMaintaining the M501NH Converter Description AutoZero Flow Check M201A Ammonia Analyzer Block Diagram No Span GAS NH3 Span GAS M201A Pneumatic Diagram Basic M201A Pneumatic Diagram Z/S Option M201A Pneumatic Diagram External Calibrator Option ECO M501NH Converter Cartridge Page M201A Spare Parts List Fuse O3, 1A M201A Spare Parts ListM201A Spare Parts List

201A specifications

The Teledyne 201A is a high-performance, compact data acquisition system that integrates advanced features and reliable technologies, making it a valuable tool for various applications in engineering, research, and industrial monitoring. This instrument is designed to deliver accurate measurements while being user-friendly and versatile.

One of the standout features of the Teledyne 201A is its modular architecture. This allows users to customize the system according to their specific needs by selecting from a range of different modules, including analog input, digital input/output, and communication interfaces. This flexibility enables users to adapt the 201A for various tasks, from simple data logging to complex multi-channel analysis.

The Teledyne 201A offers robust data acquisition capabilities with high-resolution analog-to-digital converters. This feature ensures that the system can capture even the slightest variations in data with precision, which is essential in applications such as scientific research and quality control in manufacturing processes. Additionally, it operates with a wide variety of input signal ranges, accommodating different types of sensors and devices seamlessly.

Advanced signal processing technologies are another hallmark of the Teledyne 201A. Users benefit from built-in filtering and signal conditioning, which enhances the quality of the data being collected. This is crucial for eliminating noise and improving the signal-to-noise ratio, thus ensuring that the collected data reflects true system performance.

The system's user interface is intuitive, featuring a clear display and straightforward controls, making it accessible for users at varying levels of expertise. Furthermore, it includes software that enables easy configuration and data visualization. This software provides powerful tools for real-time monitoring, analysis, and reporting, making it easier for users to interpret data and derive actionable insights.

The Teledyne 201A is designed with portability in mind. Its compact size and lightweight construction make it suitable for field deployments, allowing users to conduct measurements and monitoring in diverse environments. Whether in a laboratory, industrial setting, or outdoor location, the 201A adapts well without compromising performance.

In terms of connectivity, the Teledyne 201A supports various communication protocols, including USB, Ethernet, and serial interfaces. This ensures easy integration with existing systems and facilitates data transfer to computers or cloud-based platforms for further analysis.

In conclusion, the Teledyne 201A data acquisition system stands out due to its modular design, high-resolution data capabilities, advanced signal processing, user-friendly interface, portability, and versatile connectivity options. These features come together to make the 201A a powerful and adaptable tool for a wide range of data acquisition applications.