ResMed H4i manual Troubleshooting

Page 10

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not open the unit.

WARNING

!Service and repair should only be performed by an authorized service agent.

Problem

Possible Cause

Solution

 

 

 

Device not working

Power cord is not plugged in

(no display on LCD

properly.

screen and/or

 

indicator light not

 

lighting up)

Faulty power outlet.

 

Flow generator and docking station

 

not connected correctly.

Check that the power cord is plugged firmly into the power outlet and the switch (if available) is on at the outlet.

Try another power outlet.

Check that the docking station is securely attached to the flow generator.

Indicator light is

H4i reflector is dirty.

permanently dull and/

 

or air does not feel

 

warm

 

Clean the reflector with a damp cloth.

Leaking water

The water chamber may be

chamber

damaged or cracked.

Contact your supplier for a replacement.

No/low air flow

The air tubing is not connected

 

properly.

 

The air tubing is blocked, pinched

 

or punctured.

 

The flow generator is not working

 

correctly.

 

The cover is not latched correctly.

 

The cover seal is not fitted

 

correctly.

Connect the air tubing as described in “Setting up your H4i for use” on page 1.

Unblock or free the air tubing. Check the air tubing for punctures.

See the Troubleshooting section in the flow generator manual.

Make sure the cover is latched onto the docking station correctly.

Make sure the cover seal is facing the right way up and fitted securely.

Water splashing on

The water chamber is overfilled.

Empty some water from the water chamber.

your face

 

 

 

Condensation is forming in the air

Turn the control dial down.

 

tube and mask.

 

 

 

Use the ResMed Tubing Wrap.

 

 

 

8

Image 10
Contents H4i Contents Introduction Setting up your H4i for useComponents of your H4i System Setup Filling the Water Chamber Detaching your H4iStarting your treatment Starting and Stopping TreatmentWarming your H4i Stopping your treatmentCleaning the H4i Cleaning and MaintenanceDisassembling the H4i Cover and Cover Seal Reassembling the H4i Frequently Asked QuestionsServicing Troubleshooting Symbols which appear on the device Technical SpecificationsHumidifier Performance General Warnings Limited Warranty