Any alarm will cause the backlight to be switched on automatically.
Latching Alarms (default)
In the latched condition, once an alarm occurs, both audible and visual alarms continue to operate even after atmospheric hazard has cleared. Pressing any of the instrument buttons will clear an alarm. Any subsequent alarm will reactivate the audible and visual alarms.
In this mode, should a gas alarm occur, the instrument would enter an alarm condition. When the reading returns to normal levels, the audible and visual alarms will stop.
Vibrating Alarm
Any alarm condition that activates the audible and visual alarms will also activate the
Resetting an Alarm
If an alarm condition occurs, it is possible to cancel the alarm by pressing any of the buttons, once the gas measurements have returned to a safe level. Otherwise, the instrument will remain in the alarm condition, but the audible alarm will be muted. Any subsequent alarms that occur 1 second after the previous alarm has been reset will reactivate the audible alarm.
Fault & Warning Conditions
In addition to the gas alarms, the gas detector includes a number of auxiliary alarms to safeguard proper use of the instrument. At switch on, the gas detector performs an electronic
CAUTION
As the gas detector is designed to protect from
potentially
alarm conditions must be taken seriously.
The instrument displays a warning message for situations where a fault or error has occurred but may be resolved by the user.
See the Warning Codes in the table
Warning Codes
Number | Message | Action or Reason |
|
|
|
8 | Replace batteries | Battery unchargeable. Replace the |
|
| battery pack.. |
|
|
|
9 | Cartridge expired | Contact HSS Hire. |
|
|
|
10 | Cartridge expires in nn | Contact HSS Hire. |
| days |
|
|
|
|
11 | Calibration due | Calibration is due soon. |
|
| Recalibrate or fit new cartridge. |
14 | Battery low | Recharge the battery or fit new |
|
| battery pack. |
19 | Passed install by date | Cartridge has exceeded its |
|
| storage life. Fitting cartridge will |
|
| reduce lifetime of cartridge. |
|
|
|
20 | See manual | Cartridge not being activated. |
|
| Contact HSS Hire. |
|
|
|
24 | Please recharge or | Insufficient battery voltage to |
| replace batteries | operate the Enforcer. Recharge |
|
| battery or replace the battery |
|
| pack. |
|
|
|
25 | Calibration required | Recalibrate or fit new cartridge. |
26 | See manual | Operating temperature |
|
| specification exceeded. Operate |
|
| instrument within its specification. |
|
|
|
31 | Event log nearly full. | 20% or less remaining. When full |
| Clear log to reset | the instrument will start |
|
| overwriting the old data. Contact |
|
| HSS Hire if the downloaded log is |
|
| required. |
|
|
|
32 | Gas log nearly full. | 20% or less remaining. When full |
| Clear log to reset | the instrument will start |
|
| overwriting the old data. |
|
|
|
Fault Codes
Number | Message | Action or Reason |
|
|
|
12 | Check batteries | Mixed battery types, i.e. one dry |
|
| cell, one rechargeable. Fit two of |
|
| the same type. |
|
|
|
13 | Battery empty. Check | Battery too low to operate |
| batteries | instrument. Recharge battery or fit |
|
| new battery. |
|
|
|
23 | Switch on/off to reset | Caused by an unexpected switch |
|
| off, e.g. poor or intermittent |
|
| battery contacts. Switch off |
|
| instrument and switch on again. |
51 | Calibration required | Cell is producing an excessive |
|
| negative reading. Recalibrate |
|
|
|
53 | Calibration required | Flammable sensor has been |
|
| exposed to more than 100ppm |
|
| H2S. Recalibrate |
54 | Low O2 - flam inaccurate | Insufficient for Flammable sensor |
|
| to operate accurately. Recalibrate |
|
| or fit new cartridge. |
|
|
|
104 | Switch on/off to reset | Oxygen cell fault. Recalibrate or fit |
|
| new cartridge. |
105 | Switch on/off to reset | Flammable cell fault. Possible |
|
| Flammable fuse broken. |
|
| Recalibrate or fit new cartridge. |
|
|
|
106 | Switch on/off to reset | Toxic 1 cell fault.. Recalibrate or fit |
|
| new cartridge. |
|
|
|
107 | Switch on/off to reset | Toxic 1 cell fault.. Recalibrate or fit |
|
| new cartridge. |
TECHNICAL SPECIFICATIONS
Sensors: | Range | Repeatability |
| Response | ||
|
|
|
|
|
| time T90 |
Methane |
| 0 to | 100% LEL | ± 3% LEL |
| <15s |
|
|
|
|
|
| |
Oxygen |
| 0 to 25%v/v | ± 0.3%v/v |
| 15s | |
|
|
|
|
|
|
|
Carbon monoxide | 0 to | 500 ppm | ± 12.5 ppm |
| 25s | |
|
|
|
|
|
|
|
Hydrogen sulphide | 0 to | 50 ppm | ± 2.5 ppm |
| 25s | |
|
|
|
|
|
| |
Visual alarm | 4 |
|
|
| ||
| 2 | |||||
|
|
|
| |||
Audible alarm | 85dBA at 1m | (90dBA at 1ft) |
| |||
|
|
| ||||
Display | Large backlit graphical liquid crystal display |
| ||||
IP Rating | Instrument IP67 (NEMA 4X), cartridge IP54 (NEMA 4) | |||||
|
|
|
|
| ||
Operating |
|
|
| |||
Temperature |
|
|
|
|
|
|
EMC approvals | EN50270 |
|
|
|
| |
|
| |||||
Battery | NiMH rechargeable, operating life > 10 hours | |||||
| Charge Time = 7 Hours |
|
|
| ||
| Dry cell disposable, operating life > 17 hours | |||||
|
|
|
|
|
|
|
MAINTAINING POWER SOURCES
Portable Four Gas Detector is supplied with rechargeable battery. To charge the battery, ensure that the basestation is connected to a suitable power
source. Place the instrument in the basestation. The basestation employs a locking system to ensure that the gas detector is retained under most operating
conditions. To ensure this operates correctly, ensure that the gas detector is orientated such that the instrument lip will be inserted
under the locking lip and
rear locking catch. Whilst charging, the instrument will flash two red LEDs every 10 seconds. When charging is complete it will light the green LEDs
constantly. A fully discharged battery will typically require 7 hours to recharge.
To remove the Gas Detector press down on the rear locking catch.
Ensure the DC power supply adaptor is connected to the ‘IN’ socket on the underside of the basestation.
WARNING
Do not charge the battery pack in a hazardous area
EQUIPMENT CARE
Never push the equipment beyond its design limits.
If it is unsuitable for the task you are performing DO NOT USE OR RELY ON IT. Contact your local HSS Hire for advice.
Handle the equipment with care. Avoid dropping, knocking or otherwise exposing it to damage.
If the unit is exposed to potential contaminants such as silicone, leaded petrol, halogens, antimony or oxygen levels above 21%, return it to your local branch of HSS Hire for
Keep the equipment clean. You will find this less of a chore if you clean up regularly, rather than wait until the end of the hire period. Only use a cloth dampened with water to clean the instrument.
When not in use, store the equipment in its carrycase, somewhere clean, dry and safe from thieves.
FINISHING OFF
To turn the instrument off, press and hold the (red) button for three seconds, until it switches off.
Make sure all components are present. Give the unit a thorough clean before replacing it with the charger and line in its
... have you been trained
The law requires that personnel entering Confined Spaces using gas detectors must be competent and qualified to do so. Gas Detector and Confined Space training available at HSS Training Solutions.
0845 766 7799
…any comments?
If you have any suggestions to
enable us to improve the information within this guide please
©HSS Hire Service Group Ltd 2007 No. SS041/01 Group Office: 25 Willow Lane, Mitcham, Surrey CR4 4TS
Web Site: http://www.hss.com
SS041/01
Operating & Safety Guide SS041
Personal
Gas Detector
Continuously monitoring the
atmosphere for dangerous levels of Oxygen, Methane, Hydrogen Sulphide and Carbon Monoxide, this gas detector features both audible and visual alarms.
Code 34078