Bacharach 24-9351 Calibrate CO, For PCA Models 20, 25, 30, 35, 50, 55, 60 Material Required, 492

Page 57

PCA

Calibration

5.9 Calibrate CO

(For PCA Models 20, 25, 30, 35, 50, 55, 60 & 65)

Material Required:

Calibration Gas Fixture (Section 5.2)

Gas Cylinder, 500 ppm CO in air (Refer to Section 8.2)

Gas Cylinder, CO (1000 ppm) and H2 (1000 ppm) in Nitrogen (Refer to Section 8.2)

Procedure:

1.Enter the Calibrate Menu Screen per Section 5.3. Then choose to calibrate the CO sensor.

2.At the conclusion of Step 1 the pump should start running.

3.Using the Gas Fixture shown in Figure 5-1, attach the CO calibration- gas cylinder to the analyzer’s GAS inlet.

4.Adjust the regulator of the calibration fixture for a flowmeter indica- tion of approximately 2 SCFH.

5.Wait until the MEASURED reading on the screen stabilizes (approxi- mately 3 minutes). Then use the st and keys to enter an APPLIED value* that equals the concentration which is stamped on the CO calibration-gas cylinder.

Typical CALIBRATE CO Screen During Calibration Procedure, Using 500 ppm CO Calibration Gas:

CALIBRATE CO

 

MEASURED

492

PPM

APPLIED

0500

PPM

 

 

 

6.Press the key to calibrate the analyzer’s MEASURED value to that of the APPLIED value. At this time the Test Gas CO/H2 Screen is displayed.

*The calibration range for this screen is 250 – 1500 ppm. Any attempt to calibrate outside this range will cause the analyzer to display an error message.

Instruction 24-9351

5-9

Image 57
Contents Portable Combustion Analyzer PCA Register Your Warranty by Visiting Contents Troubleshooting Portable Combustion Analyzer IntroductionLPG Displayed DataSensor Configurations Advanced PCA Model Features PCA 35 & 65 with Draft, CO & NXPrintout Capability Technical Characteristics Normal Operating Conditions Display Agency Approvals Scope PowerUsing the Optional Power Supply Connecting the ProbeConnecting the Probe to the Analyzer Function Parameters To Change Refer to Configuring the PCAKey Pad Functions OperationSampling Hole Location Sampling Hole LocationAnalyzer Turn On and Warm Up Combustion TestInstalling Probe in the Stack Performing a Combustion Test Purging CO Sensor Turning Off the Analyzer and Purging the CO SensorDraft Differential Pressure Measurement25 MB HOT SpotBACHARACH, INC Warmup ScreenPCA Warmup yy Front Panel Key FunctionsBACHARACH, INC. PCA Warmup 0 z Sensor Status ScreenPCA models 10-25 PCA models 30, 35, 60 Combustion Test Screen«NATGAS Fuel Koks OIL NO.2 LEG OIL NO.6 LPG Coal Fuel Selection ScreenDraft Screens ‘Standard’ PCA Screen Memory Directory ScreenM9 Memory Empty Clear Memory ‘Advanced’ PCA ScreenMemory to PC Screen For ‘Advanced’ PCA Models 40, 45, 50, 55, 60Memory to PC «TRANSMIT Data Clear Memory Clear Memory COMMA-DELIMITED Fields ID Setup Screens SetupFront Panel Key Functions for the Initial ID Setup Screen PCA Setup Temperature Unit «C F Temperature Setup ScreenSetup Draft Unit «MB Draft Unit Setup ScreenLanguage Setup Screen Display Mode Setup Screen Setup Display «CO NX CU NUFor PCA Models 30, 35, 60 «TIME 154506 Date Time/Date Setup ScreenPrinter Setup Screen Setup «IR HP Printer IR IrdaRS232 Maintenance «PASSWORD Maintenance Password ScreenMaintenance «CALIBRATION User Name Maintenance ScreenUser Name «LINE 1 Line 2 Line User Name ScreensFront Panel Key Functions for Initial User Name Screen User Name LineFront Panel Key Functions for Individual User Name Screens Saving Test DataHLD 20.0 190Printing Test Data When using an infrared printer12 HLD 190 C «PRS-232Connector When using a serial printerClear Memory Screen Resetting the MicroprocessorPCA Sensor Check CalibrationProcedure Calibration FixturesMaterial Required «TA-ZERO Calibrate TA-SPAN NX TL-ZERO CO TL-SPAN Draft Calibrate Menu ScreenCalibrate TA-Zero Calibrate TA-ZERO Measured 3.0CApplied 0000.0C Calibrate TA-Span Calibrate TA-SPAN Measured 295.0CApplied 0300.0C Calibrate TL-ZERO Measured 3.0C Calibrate TL-ZeroCalibrate TL-Span Calibrate TL-SPAN Measured 102.0CApplied 0100.0C Calibrate NX Measured Calibrate NXPPM AppliedFor PCA Models 20, 25, 30, 35, 50, 55, 60 Material Required Calibrate COCalibrate CO Measured 492Typical Calibrate H2 Screen During Calibration Procedure Typical Test Gas CO/H2 Screen During Calibration ProcedureTest GAS CO/H2 CO-VALUE Enter CO-VALUE Calibrate H2 MeasuredCalibrate Draft Typical Calibrate Draft Screen During Calibration ProcedureFor PCA Models 15, 25, 35, 45, 55 Material Required Calibrate Draft Measured AppliedPCA Routine Maintenance MaintenanceDisassembling the Analyzer DisassemblyRear View of the PCB and Sensors Equipment Required Cleaning the ProbeDrain the water condensate after every use Water Trap/Filter Assembly MaintenanceReplacing the Particulate Filter Replacing the Oxygen Sensor Replacing the Nitric Oxide Sensor Filter Replacing the Nitric Oxide SensorReplacing the Nitric Oxide Sensor Bias Battery For PCA Models 20, 25, 30, 35, 50, 55, 60 Replacing the Carbon Monoxide SensorReplacing the Carbon Monoxide Sensor Filter Replacing the Pump Assembly Analyzer Repair TroubleshootingNX Sensor Error Error CodesTroubleshooting Guide Troubleshooting GuideTA-SENSOR or TL-SENSOR Troubleshooting Guide Troubleshooting Guide Replacement Parts Parts & ServiceStandard Accessories AccessoriesReplacement Parts RED Black Replacement Parts Indiana Sales/Service CentersAppendix a ENG DEU DAN FRA ESP English German Danish French SpanishDutch Swedish Finnish PolishDraft Memory Screen Polish Italian Dutch Swedish Finnish Time/Date Setup Screen Aika Calibrate TL-Span Screen Immettere Valore CO Geef DE Waarde Enter CO-VALUE PCA Appendix B Printout Translations FRA ESP POL ITA NED SVE FIN PCA World Headquarters Hunt Valley Circle, New Kensington, PA

24-9351 specifications

The Bacharach 24-9351 is a cutting-edge combustion analyzer designed for professionals in the HVAC and building maintenance industries. Known for its reliability and precision, this instrument plays a crucial role in ensuring that heating systems operate safely and efficiently.

One of the primary features of the Bacharach 24-9351 is its ability to accurately measure various combustion gases, including carbon monoxide, carbon dioxide, and oxygen levels. This multi-gas measurement capability allows technicians to diagnose combustion problems quickly and effectively, leading to improved system performance and enhanced safety for end users.

The analyzer is equipped with a high-resolution display that provides real-time readings, making it easy for technicians to analyze data on-site. Additionally, the intuitive interface is designed for ease of use, allowing users to navigate through functions and settings without extensive training.

Durability is a key characteristic of the Bacharach 24-9351. It is built to withstand the rigors of fieldwork, with a robust housing that protects it from dust, moisture, and other challenging environmental conditions. This durability is complemented by long battery life, ensuring that users can rely on the device for extended periods without interruption.

In terms of technologies, the Bacharach 24-9351 utilizes advanced sensors that ensure accuracy and repeatability in measurements. Its innovative combustion efficiency calculation provides users with essential insights into the efficiency of their heating systems, allowing for timely adjustments and tuning. This capability not only optimizes performance but also contributes to energy savings and lower operational costs.

Another notable feature is the data logging function, which allows users to store and retrieve measurement data for later analysis. This is particularly beneficial for monitoring system performance over time and making informed decisions regarding maintenance and repairs.

In summary, the Bacharach 24-9351 is an essential tool for HVAC professionals, combining accurate gas measurement, user-friendly design, and robust durability. Its advanced technologies and features make it a valuable asset for ensuring safe, efficient, and effective combustion analysis in various applications. Whether used for routine maintenance or troubleshooting, this analyzer stands out for its reliability and precision.