Fluke 566, 568 user manual Field of View, Emissivity

Page 22

566/568

Users Manual

Field of View

For accurate measurements, make sure that the target is larger than the unit’s spot size. The smaller the target, the closer you should be to it (see Figure 6).

Yes

No

 

eyl05.eps

Figure 6. Field of View

Emissivity

Emissivity describes the energy-emitting characteristics of materials. Most organic materials and painted or oxidized surfaces have an emissivity of approximately 0.95, the default setting for the thermometer.

To compensate for inaccurate readings that may result from measuring shiny metal surfaces, you can cover the surface to be measured with electrical tape or flat black paint (<148 °C/300 °F) with emissivity set to 0.95. Allow time for the tape or paint to reach the same temperature as the surface beneath it. Measure the temperature of the tape or painted surface.

If you cannot paint or use tape, then you can improve the accuracy of your measurements by either numerically adjusting emissivity or by using the Emissivity Menu to access a table of some common materials. The thermometers have numerically adjustable emissivity from 0.10 to 1.00 that can be used in conjunction with the emissivities in Table 3. Also, the thermometers have a built-in table of the most common materials that can be set from the Emissivity Menu. The built-in values are indicated in Table 3.

HOLD

The display retains its last infrared measurement for twenty seconds when the trigger is released, with HOLD appearing on the display. With the probe inserted, the contact thermometer remains on. To freeze the infrared temperature when a probe is not inserted, release the trigger until HOLD appears on the display.

14

Image 22
Contents 566/568 Limited Warranty and Limitation of Liability Table of Contents 566/568 List of Tables 566/568 List of Figures 566/568 Introduction Contacting FlukeSafety Information WWarningSymbols and Markings Symbols ExplanationFeatures Display LanguageSave Menu OverviewDescription Memory LightEmissivity Menu Nominal Surface Emissivity Material Value Min, Max, Avg, DifferentialMaterial Value Trigger Lock AlarmLaser Setup BacklightLanguage Deleting DataHow the Thermometers Work Temperature MeasurementOperating the Thermometer Locating a Hot or Cold Spot Locating a Hot or Cold SpotDistance and Spot Size Distance and Spot SizeField of View EmissivityStoring Data Downloading DataTroubleshooting Symptom Cause Action TroubleshootingExternal Contact Probe Battery Charge MaintenanceChanging the Batteries Cleaning the Housing User Replaceable Parts and AccessoriesCleaning the Lens AccessoriesRecommended Temperature Probes Specifications CE Certification Feature 566 568 Battery LifeFeature Type Thermocouple Probe Bead Type Measurement Range Cable Length566/568