Emerson Process Management 8700 manual Preparing the Transmitter

Page 5

Manual Supplement

00809-0300-4727, Rev AA June 2005

Rosemount 8700 Series

STEP 3: PREPARING THE TRANSMITTER

1.Position the remote mount terminal box adjacent to the transmitter and carefully feed the wire leads so that they are accessible from the electronics access of the transmitter.

2.Place the O-Ring in the groove on the remote mount terminal box.

3.Mount the remote mount terminal box to the transmitter using the bolts provided. See Figure 3.

4.Connect the wires from the remote mount terminal box to the electronics. Note that the sockets are different sizes so that they can only be connected one way. See Figure 4.

5.Replace the LOI (if applicable).

6.Replace the electronics side cover.

Figure 3.

Remote mount terminal box cover

Bolts

Remote mount terminal box

O-rings

Wiring

Connection

8732/8732_03a.eps

5

Image 5
Contents Integral Mount to Remote Mount Conversion Important Notice Disassembling the Tube and Transmitter Mount the junction box Preparing the FlowtubePreparing the Transmitter Wiring the Tube and Transmitter