Sanyo MBR-1404G START-UP of Unit, Handling of containers, Air intake vent, Air exhaust vent

Page 17

START-UP OF UNIT

Follow the procedures for the initial and consequent operations of the unit.

1.Connect unit to dedicated power supply. Do not put any product in the unit at this time.

2.When the power is initially applied, the audible alarm may sound. This is normal. Press the BUZZER key to silence the alarm. The audible alarm will operate until the temperature monitor bottle sensor goes into the range of 4oC±2oC.

3.Ensure that both monitor bottles have been filled with 10% glycerol as outlined earlier in this manual.

4.The unit has been factory preset to operate at 4oC±1.5oC. There is no need to set the temperature.

5.Proper chamber temperature may take several hours to stabilize at the set point temperature. Once the unit has stabilized, ensure that the monitor bottle temperature agrees with the set point temperature.

6.Turn on the light switch to ensure that the chamber light is properly functional.

7.When you are satisfied that the unit is working correctly, begin slowly placing product into the chamber to minimize the temperature rise.

Handling of containers

Arrange containers so as not to disturb air circulation in the unit. Disturbance of air circulation may cause freezing of contents or increase of temperature.

1. Air intake vent

Avoid the disturbance of air circulation in the unit by containers. Disturbance of air circulation may cause freezing of contents or rise in temperature.

2. Air exhaust vent

Do not block the exhaust air port by putting containers near the exhaust air port. Containers near the exhaust air port may freeze.

3. Load line

This unit determines the range of the containers to ensure proper air circulation. Put containers in range of the label of "Load line" which is attached on both sides inside the unit.

Fan

Air intake

ventAir exhaust vent

Load line

Air circulation

Range of the containers

Load line

Containers

16

Image 17
Contents MBR-1404G MBR-1404GR Contents Symbol shows Precautions for Safe OperationItems are stored Prevent electric shock or injury Parts are located in the upper part of this unit May result Page Environmental Conditions Refrigerator Components Inner door Upper monitor bottle Lower monitor bottle Inside of front cover Remote alarm terminal Refer to3 4 5 Control panelInstallation site InstallationGround earth InstallationFix the unit Preparation of monitor bottle Remote Alarm TerminalBefore Commencing Operation Handling of containers START-UP of UnitAir intake vent Air exhaust ventDefrost cycles Temperature displayOperating Instructions Alarm set procedure Alarm, safety function, and self diagnostic functionAlarm and safety functions Cleaning of condenser filter Routine MaintenanceCleaning of cabinet Replacement of fluorescent light Cleaning of evaporating trayCalibration of temperature sensor If nothing operates even when switched on TroubleshootingWhen unit does not get cold enough Alarm device is activatedPerformance SpecificationsSafety check sheet 7FB6P101236004 5 April

MBR-1404G, MBR-1404GR specifications

The Sanyo MBR-1404GR and MBR-1404G are high-performance, compact solar-powered refrigerators designed to cater to off-grid and energy-conscious consumers. Known for their reliability and innovative technology, these models are ideal for outdoor enthusiasts, campers, and those in remote areas requiring efficient cooling solutions.

One of the standout features of the Sanyo MBR-1404 series is its energy-saving design. These refrigerators use advanced insulation technology, allowing them to maintain optimal cooling while consuming minimal power. This makes them particularly suitable for solar power systems, where users seek to maximize energy efficiency. The MBR-1404GR and MBR-1404G can operate on low voltage, making them compatible with a wide range of renewable energy setups.

In terms of capacity, the MBR-1404GR and MBR-1404G offer ample storage space, accommodating a variety of food and beverage items. With a total capacity of approximately 140 liters, these refrigerators feature adjustable shelves that allow for customized organization, catering to different storage needs. Furthermore, the interior layout is designed to enhance airflow, ensuring consistent cooling performance throughout the unit.

The build quality of the Sanyo MBR-1404 series is robust, with a durable exterior designed to withstand harsh environmental conditions. The insulated design not only aids in energy efficiency but also enhances the unit's longevity, making it a wise investment for outdoor and off-grid living.

Both models are equipped with a user-friendly temperature control system, providing users the ability to adjust settings based on individual requirements. This level of customization is essential, especially in varying climates, allowing for optimal performance year-round.

In addition to their cooling capabilities, the MBR-1404GR and MBR-1404G are designed with silent operation in mind. The absence of noisy compressors makes these refrigerators suitable for quiet environments, ensuring a peaceful experience whether in a cabin, RV, or campsite.

Overall, the Sanyo MBR-1404GR and MBR-1404G stand out with their combination of energy efficiency, spacious interior, robust build quality, and user-friendly features. These refrigerators represent a perfect blend of technology and practicality, meeting the demands of modern consumers looking for sustainable cooling solutions without compromising on performance.