INSTALLATION SITE

To operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions:

ŶA location not subjected to direct sunlight

Do not install the unit under direct sunlight. Installation in a location subjected to direct sunlight cannot obtain the intended performance.

ŶA location with adequate ventilation

Leave at least 10 cm around the unit for ventilation. Poor ventilation will result in a reduction of the performance and consequently the failure.

ŶA location away from heat generating sources

Avoid installing the unit near heat-emitting appliances such as a heater or a boiler etc. Heat can decrease the intended performance of the unit.

ŶA location with little temperature change

Install the unit under stable ambient temperature. The allowable ambient temperature is between 5 and 30oC.

ŶA location with a sturdy and level floor

Always install the unit on a sturdy and level floor. The uneven floor or tilted installation may cause failure or injury. Install the unit in stable condition to avoid the vibration or noise. Unstable condition may cause vibration or noise.

WARNING

Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not adequate, this may result in injury from the unit falling or tipping over.

Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping. Improper installation may result in water spillage or injury from the unit tipping over.

ŶA location not prone to high humidity

Install the unit in the ambient of 80% R.H. or less humidity. Installation under high humidity may cause current leakage or electric shock.

WARNING

Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to

rain water.

Never install the unit in a humid place or a place where it is likely to be splashed by water.

Deterioration of the insulation may result which could cause current leakage or electric shock.

ŶA location without flammable or corrosive gas

Never install the unit in a flammable or volatile location. This may cause explosion or fire or may result in the current leakage or electric shock by the corrosion of the electrical components.

ŶA location without the possibility of anything fall

Avoid installing the unit in the location where anything can fall down onto the unit. This may cause the breakdown or failure of the unit.

11

44

Image 47
Contents MDF-C8V Effective models Contents Features ŶStructural specifications SpecificationsŶControl specifications ŶPerformance specifications %8 DimensionsRefrigeration circuit diagram 㪚㪐㩷 Refrigeration circuit welding pointsCooling unit parts Components on PCB Compressor terminals layout Electrical PartsSpecification of sensor Wiring diagram Circuit diagram CN3 Connection on PCBTemperature alarm Control specificationKey and Switch Temperature controlPV. Note Auto Return is not worked in F09 and F10 Other functionFunction mode F04 Purpose Simply passing through if entered by mistake Model code ‘003’ MDF-C8V Offset value Differential The point of compressor on and OFFSensor failure Remote alarm terminalWhen the power is supplied without battery Fan motor operation and compressor protection Other specificationsParts layout Electric BOX Lower back side 䇭㪥㫆䇭㫃㫆㪸㪻 Test Data䇭㪥㫆䇭㫃㫆㪸㪻 㪈㪈㪌㪭䇭㪍㪇㪟㫑䇭㪘㪫㪊㪌 䇭㪥㫆䇭㫃㫆㪸㪻 Collection and charge of refrigerant Page Page MDF-C8V Contents Introduction Precautions for Safe Operation Unqualified personnel may cause electric shock or fire Abnormal operation may cause electric shock or fire Precautions for Safe Operation Environmental Conditions Freezer Components Freezer Components Control panel Installation Site Fixing the unit InstallationAdjusting the leveling foot Removing the packaging materials and tapesSTART-UP of Unit Operation after power failureKEY Lock Function Chamber Temperature SettingAlarm Temperature Setting Alarm Temperature Setting Setting of Alarm Resume Time Change of Compressor Delay Time Remote Alarm Terminal Alarms and safety functions Alarms & Safety FunctionsDefrosting Routine MaintenanceCleaning of cabinet Battery Handling of battery Replacement of BatteryLocation of a nickel-metal-hydride battery Removal of nickel-metal-hydride batteryMalfunction Check/Remedy TroubleshootingDisposal of Unit Recycle of batteryGerman EnglishSpanish FrenchPortuguese ItalianSwedish DutchTemperature Recorder Option Installation of MTR-85H Temperature Recorder Option Installation of MTR-G85 Temperature Recorder Option BACK-UP System Option BACK-UP System Option MDF-C8V SpecificationsPerformance Safety check sheet CVK-UB4 CVK-UB4I Control Panel Trial Operation Introduction Precautions for Safe Operation Unqualified personnel may cause electric shock or fire Components List Continuation of the table Wiring ASS’YNylon Clip 6N Assembling of the solenoid valve 㪪㪺㫉㪼㫎㩷㪜㩷 Installation of the door switchInstallation of the back-up kit Installation Wiring Control Panel Calibration TemperatureTrial Operation CVK-UB4 CVK-UB4I JK1659-150B
Related manuals
Manual 2 pages 16.28 Kb

MDF-C8V specifications

The Sanyo MDF-C8V is a versatile and reliable medical refrigerator designed to meet the needs of healthcare facilities, laboratories, and pharmacies. It boasts a range of advanced features and technologies that ensure precise temperature control and safety for sensitive biological materials, vaccines, and pharmaceuticals.

One of the standout features of the Sanyo MDF-C8V is its capability to maintain a consistent temperature range between 2°C to 8°C, which is critical for the preservation of temperature-sensitive items. The unit is equipped with a microprocessor-controlled system that offers accurate temperature regulation and facilitates programmable settings. This allows users to customize the temperature according to their specific requirements.

An advanced refrigeration system is at the heart of the MDF-C8V, providing highly efficient cooling performance while minimizing energy consumption. The use of eco-friendly refrigerants ensures compliance with environmental regulations, making it an environmentally responsible choice. The cabinet is designed with high-quality insulation that enhances energy efficiency and maintains stable internal conditions.

The Sanyo MDF-C8V features an easy-to-read digital display that provides real-time temperature readings, allowing for constant monitoring of internal conditions. Additionally, an audible and visual alarm system alerts users if the temperature deviates from the preset range, enhancing the safety of stored items. This feature is particularly important in medical settings where even minor temperature fluctuations can compromise the integrity of vaccines and other critical materials.

For added convenience, the MDF-C8V comes with adjustable shelves that can accommodate various sizes of storage containers, making it easier to organize and access stored items. The unit's interior is designed to facilitate easy cleaning, ensuring hygiene standards are maintained in healthcare environments.

Furthermore, the Sanyo MDF-C8V is built with a robust exterior that is resistant to scratches and impacts, ensuring durability in busy clinical settings. Its compact design allows for easy placement in limited spaces without sacrificing storage capacity.

In summary, the Sanyo MDF-C8V is a highly functional medical refrigerator that combines precise temperature control, energy efficiency, and user-friendly features. Its advanced technologies and robust design make it an ideal solution for healthcare facilities looking to protect valuable vaccines and medications.