TEMPERATURE SETTING

Alarm temperature setting

This unit is provided with the high and low temperature alarm and the temperature at which the alarm is activated is changeable.

The following procedure shows the setting of alarm temperature according to the condition below: High temperature alarm: activates at the temperature 5oC higher than the set temperature Low temperature alarm: activates at the temperature 5oC lower than the set temperature

Note:

The alarm temperature is set at the factory 10oC higher and lower than the set temperature.

The available range of alarm temperature is between 5oC and 20oC higher or lower than the set temperature.

Table 3. Procedure for setting high temperature alarm

 

Description of operation

Key operated

1

 

----

 

 

 

2

Press numerical value shift key for

 

about 5 seconds.

 

 

 

3

Press numerical value shift key

 

and scroll the figure to 1.

 

 

 

4

Press SET key.

SET

 

 

 

 

Scroll the figure to 005 by using

 

5

digit shift key and numerical value

 

 

 

shift key

 

 

 

 

6

Press SET key.

SET

 

 

 

Indication after operation The current chamber temperature is displayed.

The first digit is flashed.

The first digit is flashed.

The first digit is flashed.

When pressed, the figure of settable digit changes.

When pressed, the changeable digit moves.

Alarm temperature is memorized and the current chamber temperature is displayed.

Table 4. Procedure for setting low temperature alarm

 

Description of operation

Key operated

1

 

----

 

 

 

2

Press numerical value shift key for

 

about 5 seconds.

 

 

 

3

Press numerical value shift key

 

and scroll the figure to 2..

 

 

 

4

Press SET key.

SET

 

 

 

 

Scroll the figure to -05 by using

 

5

digit shift key and numerical value

 

 

 

shift key

 

 

 

 

6

Press SET key.

SET

 

 

 

Indication after operation The current chamber temperature is displayed.

The first digit is flashed.

The first digit is flashed.

The first digit is flashed.

When pressed, the figure of settable digit changes.

When pressed, the changeable digit moves.

Alarm temperature is memorized and the current chamber temperature is displayed.

18

Image 20
Contents MDF-U32V Page Contents Precautions for Safe Operation Symbol showsMay cause bad effect on your health or environment Prevent accidents such as suffocation Water Page Environmental Conditions Freezer Components Space for optional component Control panel and keypad Ultra LOWInstallation Installation siteInstallation Adjust the leveling footRemove the packaging materials and tapes Fix the unitTemperature recorder Optional ComponentsBack-up system Inventory rack START-UP of Unit Key lock function Temperature SettingChamber temperature Alarm temperature setting Alarms & Safety Functions Alarms and safety functionsSetting of Alarm Resume Time Remote Alarm Terminal Monitor of Freezer Status Monitor of running statusChange of Compressor Delay Time Cleaning of condenser filter Routine MaintenanceCleaning of cabinet Defrosting of inside wall Troubleshooting Disposal of UnitRemoval of nickel-metal-hydride battery Recycle of Rechargeable BatteryLocation of a nickel-metal-hydride battery Specifications PerformanceSafety check sheet 7FB6P10134200

MDF-U32V specifications

The Sanyo MDF-U32V is a high-performance laboratory freezer designed to meet the stringent requirements for preserving sensitive samples. This model is particularly popular in medical, pharmaceutical, and research environments, where maintaining low temperatures is crucial. One of its standout features is its superior cooling performance, capable of maintaining a temperature range as low as -30 degrees Celsius. This capability ensures that biological samples, vaccines, and other temperature-sensitive items are stored safely and effectively.

Equipped with a highly efficient compressor, the MDF-U32V utilizes advanced technology to provide rapid cooling and stable temperature control. The freezer is designed with a cycle system that minimizes temperature fluctuations, offering peace of mind to users who rely on stable storage conditions. The insulation used in the MDF-U32V aids in energy conservation, reducing the overall running costs while maintaining optimal performance.

One of the significant innovations in the Sanyo MDF-U32V is its user-friendly digital control panel. This interface allows for precise temperature setting and monitoring, making it easy for lab personnel to operate. Additionally, the system features an alarm for temperature deviations, ensuring that users are alerted immediately to any changes that could compromise sample integrity.

The MDF-U32V also boasts a spacious interior with adjustable shelving, providing flexibility in sample organization. The interior is designed for easy cleaning, which is an important feature in laboratory settings where contamination risks must be minimized. Furthermore, the solid construction of the freezer is designed to withstand demanding environments, ensuring durability and longevity.

Another noteworthy characteristic is the noise level of the MDF-U32V, which operates quietly, making it suitable for use in settings where noise might be a concern. The freezer is also equipped with an automatic defrost function, reducing maintenance needs and improving reliability.

In summary, the Sanyo MDF-U32V is an exemplary choice for those looking for a dependable laboratory freezer. With its excellent cooling capabilities, advanced digital controls, durable design, and efficient performance, it stands out as a reliable solution for preserving valuable samples in a variety of professional environments. Whether for biological research, clinical applications, or pharmaceutical storage, the MDF-U32V meets the high standards necessary for scientific success.