above, press the ZERO switch to display the decimation ratio (filtering). The larger the ratio, the longer the response time.

EEPROM Calibration Restoration

There are interlocks built into the CAL mode to minimize the chance of accidental alteration of the calibration coefficients, but if calibration alteration does occur, the EEPROM calibration data can be restored in the RUN mode by pressing the UP and DOWN switches simultaneously (RUN light will flash), then press the “ZERO” switch, UP and DOWN switches (all three) simultaneously.

This will not correct the calibration if the tube has become contaminated or damaged.This reset will not affect the gas selection, units or setpoints.

Also see units mode for default calibration restoration.

3.3 HIGH and LOW Set Point Modes

The Model 2002 provides TTL outputs for process control. The I/O cable is attached via a 9-pin sub-D connector to the rear panel of the control unit. The pinout is shown in Figure 4.2.

To view the high set point, place the Model 2002 in the HIGH mode by pressing the MODE switch to illuminate the HIGH light and no other mode light. The display will show the set point selected. During normal operation the alarm light will illuminate and the TTL output (pin # 3) will go high (+5 V) if the pressure exceeds the set point.

Similarly, to view the low set point access LOW mode by pressing the mode switch until the LOW light is lit and no other mode light is illuminated. During normal operation, the LOW light will illuminate and the LOW alarm TTL output (pin # 4) will go high (+5 V) if the pressure becomes less than the set point.

The alarm lights cannot indicate an alarm condition while in the set point modes. Therefore, it is advisable not to leave the instrument in these modes for extended periods.

To adjust a set point, press the mode switch until the set point is viewed. Then simultaneously press the UP and DOWN switches. The active mode light will now flash indicating that the interlock has been bypassed and the set point can be adjusted. The display shows the present set point.

Enter the new set point by using the UP and DOWN switches. Press and hold the UP switch to increase the set point and the DOWN switch to decrease the setpoint. Allow a few seconds for the circuitry to respond. If the button is held down for and extended period of time the adjustment rate will start to increase. Release the button and press again to get finer control of the exact trigger point.

At this point, the new set point is in temporary memory. If the instrument were unplugged now, the Model 2002 would revert back to the original set point upon restarting. Return the 2002 to the RUN mode to store the set point in permanent memory.

Once the interlock has been bypassed in the HIGH mode, it will stay active until the CAL mode is entered. This will allow the user to set both setpoints without repeating the interlock bypass.

3.4 CAL Mode

Optimal performance of the Model 2002 is achieved by performing in situ adjustments to the calibration coefficient in the CAL mode. There are three calibration coefficients. These are the zero coefficient, the midrange coefficient and the atmosphere coefficient. Once a tube has been fully calibrated the midrange coefficient should never need further adjustment, but it may be helpful to adjust the zero coefficient or the atmosphere coefficient under certain situations. Even though the operator inputs may be identical for adjustment of all three coefficients, microprocessor will detect

Model 2002 page 11

Image 11
Contents Teledyne Hastings Manual Print History Table of Contents 12.0 4-20mA Interface Option BoardVolt Interface Model 2002 Sensors FeaturesGeneral Information Model 2002 Control Unit SpecificationsQuick Start InstallationReceiving Inspection Transducer Cable Attachment Transducer InstallationControl Unit Installation Environment Panel Mount InstructionsModel Control Unit Front Panel Front Panel OperationOverall Functional Description Scientific Notation RUN Mode Normal OperationSpeed Adjust High and LOW Set Point Modes CAL ModeEeprom Calibration Restoration Maintainance and Repair Midrange Coefficient Adjustment Zero Coefficient AdjustmentGAS Mode Atmosphere Coefficient AdjustmentDefault Calibration Restoration Units ModeThis page was intentionally left blank Analog Output Rear Panel DescriptionRemote Zero Input Voltage to pressure conversion example TTL Outputs Model 2002 Analog OutputModel Rear Panel Detail Power Entry ModulePiezoresistive Sensor Theory of OperationVo = SPV+V1 View A-A Er = σεTh4-Ta4As Pirani SensorEt = Es + Er + Eg Es = K ∆T Am/LPage Dual Sensor Operation Eg = Kg ∆T As/∆x= a L 273/T 1/2T -T a P Model Dual Sensor Vacuum Gauge 28 Model Communications Option Board RS-232-E Interface SpecificationsFor RS-485 Half Duplex 2 wire Interface Connector Pin Assignments for RS-485Interrogation Commands Command SyntaxCalibration Adjustment Commands Parameter Modification CommandsReset / Restore Commands Digit# Device StatusRelay Interface Option Board Relay Board SpecificationsRelay Connector Pin Assignment Operation MA Interface Option Board Channel 10V Interface Option Board 0-10V Interface Connector Pin AssignmetsVchannel 1 = PTorr 100 Vchannel 2 = PmTorr Troubleshooting Guide 42 Model Non-Warranty Repair Policy Warranty and RepairWarranty Repair Policy 44 Model Diagrams and Drawings KF-25 HPM-2002-06 Mini Conflattm HPM-2002-03KF-16 HPM-2002-05 4 Conflattm HPM-2002-04 NPT HPM-2002-01O.D. Smooth Tube HPM-2002-07 VCR HPM-2002-02Panel Cut-Out Dimensions Model Out Line DimensionsMounting Clip Attachment Panel Mounting50 Model Model 2002 52 Model Model 2002 54 Model
Related manuals
Manual 37 pages 47.27 Kb Manual 38 pages 51.68 Kb

2002 specifications

Teledyne 2002 represents a significant advancement in the realm of sophisticated instrumentation and systems used across various industries. This innovative platform emerged as a versatile solution for a multitude of applications including environmental monitoring, industrial automation, and scientific research.

One of the most notable features of the Teledyne 2002 is its robust data analysis capability. Equipped with powerful processing units, it allows users to conduct real-time data analysis, ensuring accurate and timely results. This is particularly beneficial in fields where immediate decision-making is crucial, such as environmental assessments and industrial quality control.

The technology behind the Teledyne 2002 encompasses a variety of sensors and analytical instruments. Its modular design enables users to customize the system according to their specific needs, integrating various sensors such as gas analyzers, spectrometers, and temperature sensors. This flexibility makes the Teledyne 2002 applicable in diverse settings, from laboratory environments to rugged field conditions.

Another characteristic of the Teledyne 2002 is its user-friendly interface. The system is designed with an intuitive control panel and advanced software that provides comprehensive data visualization. This ease of use enhances productivity, allowing operators to facilitate complex analyses without extensive training.

Security and data integrity are also focal points of the Teledyne 2002. The system implements state-of-the-art encryption protocols to protect sensitive data. Moreover, it maintains compliance with industry standards, ensuring that the collected data is reliable and trustworthy.

In addition to its technological features, the Teledyne 2002 is built with durability in mind. The rugged construction allows it to withstand harsh environmental conditions, making it ideal for outdoor applications. Its compact design enables easy transportation, which is essential for fieldwork.

Moreover, the Teledyne 2002 includes connectivity options that facilitate seamless integration with existing systems. It can connect to cloud services, enabling remote monitoring and data storage. This feature not only enhances the functionality of the device but also allows for collaborative data sharing among teams.

Overall, the Teledyne 2002 symbolizes a convergence of advanced technology, user-centric design, and robust performance. With its extensive features and versatility, it stands out as a premier solution for professionals demanding precision and reliability in their analytical endeavors. Whether for environmental monitoring or industrial applications, the Teledyne 2002 is equipped to meet the challenges of modern data analysis.