Set Full Scale Calibration

CF={m.d}E{e}<CR> 5.12e+2 to 1.023e+3 Torr

 

6.83e+2

to 1.365e+3 mbar

 

6.83e+4

to 1.365e+5 Pascal

Set Midpoint Calibration

CM={m.d}E{e}<CR> 4.00e+0 to 7.999e+0 Torr

 

5.34e+0

to 1.066e+1 mbar

 

5.34e+2

to 1.066e+3 Pascal

Set Lowpoint Calibration

CL={m.d}E{e}<CR>

 

 

0 to 1.249e-1 Torr

 

0 to 1.666e-1 mbar

 

0 to 1.666e+1 Pascal

Notes:

The calibration adjustment data may also be entered as a decimal number, e.g. [CF=760.99<CR>] will be same as entering [CF=7.6099E+2<CR>] .

When inputting calibration adjustment data, it must be within the valid range of the presently selected Unit of Measurement (i.e.Torr, mbar or Pascal).The data is checked to be valid before being accepted.

If the command syntax is not met, or if the number is out or range, the HPM-2002 will respond with the ASCII codes for <bell>?<CR>, and the command will be ignored.

Reset / Restore Commands

Command Description

Format

Notes :

Escape

<Esc>

Reset Command Buffer (ignore prior

Input

 

 

Software Reset

/R<CR>

Reinitialize Software

Restore Factory Defaults

/#<CR>

Restore Calibration Register Default

Values

 

 

Set Zero

/0 <CR>

Store Present Pressure As Instrument

 

 

ZERO (if Piezo <32Torr & Pirani <50

mTorr)

 

 

Force Zero

/!<CR>

Override Limit Checks And Store

Present

 

 

 

 

Pressure As Instrument ZERO

Model 2002 page 33

Image 33
Contents Teledyne Hastings Manual Print History Table of Contents 4-20mA Interface Option Board Volt Interface12.0 Features General InformationModel 2002 Sensors Model 2002 Control Unit SpecificationsInstallation Receiving InspectionQuick Start Control Unit Installation Environment Transducer InstallationPanel Mount Instructions Transducer Cable AttachmentFront Panel Operation Overall Functional DescriptionModel Control Unit Front Panel RUN Mode Normal Operation Speed AdjustScientific Notation CAL Mode Eeprom Calibration RestorationHigh and LOW Set Point Modes Maintainance and Repair Midrange Coefficient Adjustment Zero Coefficient AdjustmentGAS Mode Atmosphere Coefficient AdjustmentDefault Calibration Restoration Units ModeThis page was intentionally left blank Rear Panel Description Remote Zero InputAnalog Output Voltage to pressure conversion example TTL Outputs Model 2002 Analog OutputModel Rear Panel Detail Power Entry ModulePiezoresistive Sensor Theory of OperationVo = SPV+V1 View A-A Et = Es + Er + Eg Pirani SensorEs = K ∆T Am/L Er = σεTh4-Ta4AsPage Eg = Kg ∆T As/∆x = a L 273/T 1/2T -T a PDual Sensor Operation Model Dual Sensor Vacuum Gauge 28 Model Communications Option Board RS-232-E Interface SpecificationsFor RS-485 Half Duplex 2 wire Interface Connector Pin Assignments for RS-485Interrogation Commands Command SyntaxCalibration Adjustment Commands Parameter Modification CommandsReset / Restore Commands Digit# Device StatusRelay Interface Option Board Relay Board SpecificationsRelay Connector Pin Assignment Operation MA Interface Option Board Channel 10V Interface Option Board 0-10V Interface Connector Pin AssignmetsVchannel 1 = PTorr 100 Vchannel 2 = PmTorr Troubleshooting Guide 42 Model Warranty and Repair Warranty Repair PolicyNon-Warranty Repair Policy 44 Model Diagrams and Drawings Mini Conflattm HPM-2002-03 KF-16 HPM-2002-05KF-25 HPM-2002-06 O.D. Smooth Tube HPM-2002-07 NPT HPM-2002-01VCR HPM-2002-02 4 Conflattm HPM-2002-04Panel Cut-Out Dimensions Model Out Line DimensionsMounting Clip Attachment Panel Mounting50 Model Model 2002 52 Model Model 2002 54 Model
Related manuals
Manual 37 pages 47.27 Kb Manual 38 pages 51.68 Kb

2002 specifications

Teledyne 2002 represents a significant advancement in the realm of sophisticated instrumentation and systems used across various industries. This innovative platform emerged as a versatile solution for a multitude of applications including environmental monitoring, industrial automation, and scientific research.

One of the most notable features of the Teledyne 2002 is its robust data analysis capability. Equipped with powerful processing units, it allows users to conduct real-time data analysis, ensuring accurate and timely results. This is particularly beneficial in fields where immediate decision-making is crucial, such as environmental assessments and industrial quality control.

The technology behind the Teledyne 2002 encompasses a variety of sensors and analytical instruments. Its modular design enables users to customize the system according to their specific needs, integrating various sensors such as gas analyzers, spectrometers, and temperature sensors. This flexibility makes the Teledyne 2002 applicable in diverse settings, from laboratory environments to rugged field conditions.

Another characteristic of the Teledyne 2002 is its user-friendly interface. The system is designed with an intuitive control panel and advanced software that provides comprehensive data visualization. This ease of use enhances productivity, allowing operators to facilitate complex analyses without extensive training.

Security and data integrity are also focal points of the Teledyne 2002. The system implements state-of-the-art encryption protocols to protect sensitive data. Moreover, it maintains compliance with industry standards, ensuring that the collected data is reliable and trustworthy.

In addition to its technological features, the Teledyne 2002 is built with durability in mind. The rugged construction allows it to withstand harsh environmental conditions, making it ideal for outdoor applications. Its compact design enables easy transportation, which is essential for fieldwork.

Moreover, the Teledyne 2002 includes connectivity options that facilitate seamless integration with existing systems. It can connect to cloud services, enabling remote monitoring and data storage. This feature not only enhances the functionality of the device but also allows for collaborative data sharing among teams.

Overall, the Teledyne 2002 symbolizes a convergence of advanced technology, user-centric design, and robust performance. With its extensive features and versatility, it stands out as a premier solution for professionals demanding precision and reliability in their analytical endeavors. Whether for environmental monitoring or industrial applications, the Teledyne 2002 is equipped to meet the challenges of modern data analysis.