Both RS-232 and RS-485 communication interface boards are available as an option for the Model 2002.The communication option boards allow data to be output to a host computer with the appropiate interface. Connection to the RS-232 or RS-485 communication interface boards are via a 9 pin D-subminiature female connector.

6.1RS-232-E Interface Specifications

Format

EIA standard RS-232-E, full-duplex, no handshaking, asynchronous

Data Rate

9600 baud

Character Length

Eight data bits

Parity

No parity

Stop bits

One stop bit

A common application of the RS232 version of the HPM-2002, is to connect the pressure gauge directly to the serial port of a PC. This is done by first wiring a communication cable in the

Figure 6.1

manner shown below.

PIN

MNEMONIC

SIGNAL

DESCRIPTION

 

 

 

 

 

 

 

 

2

TX

Transmit Data

Transmits data within RS-232-E voltage levels

 

 

3

RX

Receive Data

Accepts data within RS-232-E voltage levels.

 

 

5

SG

Signal Ground

Establishes the common ground reference

 

 

 

 

 

potential for all interchange circuits

 

 

when

 

 

communicating via RS-

 

 

232-E.

 

 

 

 

 

4

CTS

Clear to send

Host computer ready to accept data. (Reserved

 

 

 

 

 

for future use, not implemented by present

 

 

 

 

 

software.

 

 

8

RDY

Ready

Ready to receive data (always high, not

 

 

 

 

 

implemented by present software.

 

 

 

 

 

 

 

 

 

Model 2002 page 29

Image 29
Contents Teledyne Hastings Manual Print History Table of Contents 12.0 4-20mA Interface Option BoardVolt Interface Model 2002 Sensors FeaturesGeneral Information Model 2002 Control Unit SpecificationsQuick Start InstallationReceiving Inspection Control Unit Installation Environment Transducer InstallationPanel Mount Instructions Transducer Cable AttachmentModel Control Unit Front Panel Front Panel OperationOverall Functional Description Scientific Notation RUN Mode Normal OperationSpeed Adjust High and LOW Set Point Modes CAL ModeEeprom Calibration Restoration Maintainance and Repair Midrange Coefficient Adjustment Zero Coefficient AdjustmentGAS Mode Atmosphere Coefficient AdjustmentDefault Calibration Restoration Units ModeThis page was intentionally left blank Analog Output Rear Panel DescriptionRemote Zero Input Voltage to pressure conversion example TTL Outputs Model 2002 Analog OutputModel Rear Panel Detail Power Entry ModulePiezoresistive Sensor Theory of OperationVo = SPV+V1 View A-A Et = Es + Er + Eg Pirani SensorEs = K ∆T Am/L Er = σεTh4-Ta4AsPage Dual Sensor Operation Eg = Kg ∆T As/∆x= a L 273/T 1/2T -T a P Model Dual Sensor Vacuum Gauge 28 Model Communications Option Board RS-232-E Interface SpecificationsFor RS-485 Half Duplex 2 wire Interface Connector Pin Assignments for RS-485Interrogation Commands Command SyntaxCalibration Adjustment Commands Parameter Modification CommandsReset / Restore Commands Digit# Device StatusRelay Interface Option Board Relay Board SpecificationsRelay Connector Pin Assignment Operation MA Interface Option Board Channel 10V Interface Option Board 0-10V Interface Connector Pin AssignmetsVchannel 1 = PTorr 100 Vchannel 2 = PmTorr Troubleshooting Guide 42 Model Non-Warranty Repair Policy Warranty and RepairWarranty Repair Policy 44 Model Diagrams and Drawings KF-25 HPM-2002-06 Mini Conflattm HPM-2002-03KF-16 HPM-2002-05 O.D. Smooth Tube HPM-2002-07 NPT HPM-2002-01VCR HPM-2002-02 4 Conflattm HPM-2002-04Panel Cut-Out Dimensions Model Out Line DimensionsMounting Clip Attachment Panel Mounting50 Model Model 2002 52 Model Model 2002 54 Model
Related manuals
Manual 37 pages 47.27 Kb Manual 38 pages 51.68 Kb

2002 specifications

Teledyne 2002 represents a significant advancement in the realm of sophisticated instrumentation and systems used across various industries. This innovative platform emerged as a versatile solution for a multitude of applications including environmental monitoring, industrial automation, and scientific research.

One of the most notable features of the Teledyne 2002 is its robust data analysis capability. Equipped with powerful processing units, it allows users to conduct real-time data analysis, ensuring accurate and timely results. This is particularly beneficial in fields where immediate decision-making is crucial, such as environmental assessments and industrial quality control.

The technology behind the Teledyne 2002 encompasses a variety of sensors and analytical instruments. Its modular design enables users to customize the system according to their specific needs, integrating various sensors such as gas analyzers, spectrometers, and temperature sensors. This flexibility makes the Teledyne 2002 applicable in diverse settings, from laboratory environments to rugged field conditions.

Another characteristic of the Teledyne 2002 is its user-friendly interface. The system is designed with an intuitive control panel and advanced software that provides comprehensive data visualization. This ease of use enhances productivity, allowing operators to facilitate complex analyses without extensive training.

Security and data integrity are also focal points of the Teledyne 2002. The system implements state-of-the-art encryption protocols to protect sensitive data. Moreover, it maintains compliance with industry standards, ensuring that the collected data is reliable and trustworthy.

In addition to its technological features, the Teledyne 2002 is built with durability in mind. The rugged construction allows it to withstand harsh environmental conditions, making it ideal for outdoor applications. Its compact design enables easy transportation, which is essential for fieldwork.

Moreover, the Teledyne 2002 includes connectivity options that facilitate seamless integration with existing systems. It can connect to cloud services, enabling remote monitoring and data storage. This feature not only enhances the functionality of the device but also allows for collaborative data sharing among teams.

Overall, the Teledyne 2002 symbolizes a convergence of advanced technology, user-centric design, and robust performance. With its extensive features and versatility, it stands out as a premier solution for professionals demanding precision and reliability in their analytical endeavors. Whether for environmental monitoring or industrial applications, the Teledyne 2002 is equipped to meet the challenges of modern data analysis.