9.ANL unit No. | 1 | ‘1’: AU5400 ISE | ANL unit No. is added to the |
|
| ‘1’: AU5400 unit 1 | header portion of RB, ACAL, |
(Optional) |
| ‘2’: AU5400 unit 2 | and QC result data. Normal |
|
| ‘3’: AU5400 unit 3 | sample and rerun sample, it is |
|
|
| added to each item. However, |
|
| ‘1’: AU2700 | in the case of an calculate |
|
|
| item, a space is added. In a |
|
|
| normal sample and repeat run, |
|
|
| it can be prevented from |
|
|
| adding a unit number by the |
|
|
| setup of an |
|
|
| menu. This option can be |
|
|
| selected in the |
|
|
| [PARAMETER][ONLINE] |
|
|
| [PROTOCOL][SPECIAL] |
|
|
| menu. |
10.Cuvette side | 1 | AU5400 | Cuvette side is added to the |
|
| ‘0’ : inner cuvette (ISE Cell 1) | header portion of RB, ACAL, |
(Optional) |
| ‘1’ : outer cuvette (ISE Cell 2) | and QC result data. Normal |
|
| AU2700 | sample and rerun sample, it is |
|
| ‘0’ : inner cuvette(ISE) | added to each item. However, |
|
| ‘1’ : outer cuvette | in the case of an calculate |
|
|
| item, a space is added. In a |
|
|
| normal sample and rerun, it |
|
|
| can be prevented from adding |
|
|
| a unit number by the setup of |
|
|
| an |
|
|
| This option can be selected in |
|
|
| the |
|
|
| [PARAMETER][ONLINE] |
|
|
| [PROTOCOL][SPECIAL] |
|
|
| menu. |
11. Sex | 1 | M : male | • These items can be |
|
| F : female | selected/canceled in |
(Optional) |
| SP : none sex | the [PARAMETER] |
|
| 0 : | [FORMAT] |
|
|
| [FORMAT |
|
|
| REQUISITION] |
|
|
| menu. |
12. Age | 3 |
| |
(Optional) |
| space : |
|
13. Month | 2 |
| |
(Optional) |
| space : |
|
14. Patient | 20 | number or character |
|
Information |
|
|
|
(Optional) |
|
|
|
15. Online | 2 | • This number can be | |
Test No. |
|
| changed in the |
|
|
| [PARAMETER] |
|
|
| [ONLINE] [TEST |
|
|
| NO.] |
16. Method | 1 | '0':normal | For Repeat Run Results. |
|
| '1':dilution |
|
|
| ‘2’:condense |
|
- 23 -