An important point to be noted is the fact that after the adjustment, the new calibration coefficient is only in temporary memory. If the instrument were unplugged at this point, it would revert back to the original displayed value upon restarting. Place the Model 2002 back in the RUN mode to store all of the current calibration coefficients in permanent memory.

To perform a full calibration on the Model 2002 system, first use the Zero Coefficient Adjustment Procedure. Followed by the Midrange Coefficient Adjustment Procedure, and finally perform the Atmosphere Coefficient Adjustment Procedure.

Zero Coefficient Adjustment

The zero coefficients corrects for the constant power level which is present over the entire pressure range. Typically, this adjustment corrects for low pressure errors. The instrument will need to be re-zeroed often if measurements are being made in the 10-4 Torr range, especially if the ambient temperature changes. The transducer may have a temperature coefficient of up to 2x10-4 Torr/oC. The instrument remote zero input will allow an external gauge such as an ion gauge to automatically re-zero the Model 2002 whenever the pressure drops below the desired pressure level (if it has a TTL output).

NOTE: Do not attempt to zero the Model 2002 in pressures above 10-2 Torr; the microprocessor will not accept a zero above this pressure.

To manually adjust the zero uses the following procedure:

1.If possible, evacuate the vacuum system into the low 10-6 Torr (1.33x10-6 mbar) range or as low as possible below 10-4 Torr.

2.Allow the sensor to operate in this condition for a minimum of 15 minutes.

Model 2002 Vacuum Gauge

Page 11 of 37

Image 11
Contents Teledyne Hastings Manual Print History Table of Contents Model 2002 Control Unit FeaturesGeneral Information Model 2002 SensorsSpecifications Quick Start InstallationTransducer Installation Receiving InspectionTransducer Cable Attachment Control Unit InstallationInitial Operation Panel Mount InstructionsOverall Functional Description Front Panel OperationNormal Operation RUN ModeSpeed Adjust Scientific NotationHigh and Low Set Point Modes CAL ModePage Atmosphere Coefficient Adjustment GAS ModeMidrange Coefficient Adjustment Default Calibration Restoration Units ModeAnalog Output Rear Panel DescriptionRemote Zero Input Voltage to pressure conversion example Model Rear Panel Detail Power Entry ModuleTTL Outputs Dual Sensor Emodiment Theory of OperationPiezoresistive Sensor = ΔRL Et = Es + Er + Eg Pirani SensorEg = Kg DT As/Dx Dual Sensor Operation Communications Option Board RS-232-E Interface SpecificationsFor RS-485 Half Duplex 2 wire Interface Connector Pin Assignments for RS-485Parameter Modification Commands Interrogation CommandsCalibration Adjustment Commands Device Status Reset / Restore CommandsRelay Interface Option Board Relay Board SpecificationsRelay Connector Pin Assignment MA Interface Option Board 20 mA Interference Connector Pin AssignmentsOperation 10V Interface Option Board 0-10V Interface Connector Pin Assignmets= PTorr Channel1 Troubleshooting Guide Non-Warranty Repair Policy Warranty Repair PolicyDiagrams and Drawings Model 2002 Vacuum Gauge Model Outline Dimensions Model 2002 Vacuum Gauge Model 2002 Vacuum Gauge
Related manuals
Manual 54 pages 12.52 Kb Manual 38 pages 51.68 Kb

2002 specifications

Teledyne 2002 represents a significant advancement in the realm of sophisticated instrumentation and systems used across various industries. This innovative platform emerged as a versatile solution for a multitude of applications including environmental monitoring, industrial automation, and scientific research.

One of the most notable features of the Teledyne 2002 is its robust data analysis capability. Equipped with powerful processing units, it allows users to conduct real-time data analysis, ensuring accurate and timely results. This is particularly beneficial in fields where immediate decision-making is crucial, such as environmental assessments and industrial quality control.

The technology behind the Teledyne 2002 encompasses a variety of sensors and analytical instruments. Its modular design enables users to customize the system according to their specific needs, integrating various sensors such as gas analyzers, spectrometers, and temperature sensors. This flexibility makes the Teledyne 2002 applicable in diverse settings, from laboratory environments to rugged field conditions.

Another characteristic of the Teledyne 2002 is its user-friendly interface. The system is designed with an intuitive control panel and advanced software that provides comprehensive data visualization. This ease of use enhances productivity, allowing operators to facilitate complex analyses without extensive training.

Security and data integrity are also focal points of the Teledyne 2002. The system implements state-of-the-art encryption protocols to protect sensitive data. Moreover, it maintains compliance with industry standards, ensuring that the collected data is reliable and trustworthy.

In addition to its technological features, the Teledyne 2002 is built with durability in mind. The rugged construction allows it to withstand harsh environmental conditions, making it ideal for outdoor applications. Its compact design enables easy transportation, which is essential for fieldwork.

Moreover, the Teledyne 2002 includes connectivity options that facilitate seamless integration with existing systems. It can connect to cloud services, enabling remote monitoring and data storage. This feature not only enhances the functionality of the device but also allows for collaborative data sharing among teams.

Overall, the Teledyne 2002 symbolizes a convergence of advanced technology, user-centric design, and robust performance. With its extensive features and versatility, it stands out as a premier solution for professionals demanding precision and reliability in their analytical endeavors. Whether for environmental monitoring or industrial applications, the Teledyne 2002 is equipped to meet the challenges of modern data analysis.