HV-WP/HW-WP Series Page 9 5. 1. Removing the ploe
5.1.
5.1.5.1.
5.1. Removing the
Removing the Removing the
Removing the pole
polepole
pole
Caution
Remove the main power plug from the receptacle before removing the pole.
When removing the loadcell cable, do not pull the loadcell cable connector
forcibly and do not pull on the wires of the cable.
Do not bend the cable forcibly.
Avoid dust, static electricity and high humidity (or drops) because the inside
of the indicator unit is very sensitive.
Procedure
Step 1 Remove the power plug from the receptacle.
Step 2 Open the rear cover of indicator unit.
Disconnect the loadcell cable connector gently
(perpendicularly, do not pull to the side).
Step 3 Remove the ferrite core and cable clamp from
the loadcell cable.
Step 4 Loosen the knobs to remove the indicator unit.
Step 5 Remove four 3mm screws from the bottom of the
pole base for HV-60KV-WP, HV-200KV-WP,
HW-60KV-WP, HW-100KV-WP, HW-200KV-WP.
Step 6 Carefully remove the cable from the pole and
pole base. Especially, use care with the HV-
15KV-WP, HW-10KV-WP so that the connector
is not pulled forcibly.
Step 7 Arrange the cable so that it does not touch to
the weighing pan in the base unit. The untied
cable is at least 2m long. The optional
extension loadcell cable (OP-02) is 5m long.