TROUBLESHOOTING

 

 

 

 

Corrective action

 

Problem

 

Probable cause

 

Display is blank,

 

Battery is drained.

Replace with a new battery.

 

even when the

 

Battery terminals are

Place the battery with

 

power is turned

 

 

on.

 

not in the correct

negative and positive

 

 

 

position.

terminals matching those

 

 

 

 

indicated in the battery

 

 

 

 

compartment.

 

The device will

 

The cuff is not

Fasten the cuff correctly.

 

not measure.

 

fastened properly.

 

 

Readings are too

 

You moved your arm

Make sure you remain very

 

high or too low.

 

 

 

 

or body during the

still and quiet during the

 

 

 

measurement.

measurement.

 

 

 

The cuff position is

Adjust the cuff position.

 

 

 

not correct.

Raise your hand so that the

 

 

 

 

cuff is at the same level as

 

 

 

 

your heart.

 

 

 

You are wearing the

See Pg. E-5 “Selecting the

 

 

 

wrong cuff size.

Correct Cuff.”

 

Other

 

The blood pressure

See Pg. E-12 “Keys to

 

 

 

reading is different

Successful Monitoring.”

 

 

 

from that measured

Remove the battery.

 

 

 

at a clinic or doctor’s

 

 

 

office.

Place back in correctly, and

 

 

 

 

try the measurement again.

 

 

 

Cuff will not inflate.

Check air tube connection

 

 

 

 

inside battery cover.

 

 

 

 

 

 

 

 

 

 

NOTE: If the actions described above do not solve the problem, contact A&D Medical or Auto Control Medical at the numbers listed at the bottom of these pages. Do not attempt to repair the device yourself.

E-14

LifeSource Health Line (Toll-Free): 1-888-726-9966

 

SPECIFICATIONS

 

 

 

Model

. . . .UA-704

 

Type

. . . .Oscillometric

 

Display

. . . .Digital, 10-mm character height

 

Pressure/pulse displayed

 

Measurement range . . . .

alternately

 

. . . .Pressure: 20 mmHg to 280 mmHg

 

Pulse: 40 pulses to 200

 

Accuracy

pulses/minute

 

. . . .Pressure: +3 mmHg or 2%,

 

whichever is greater

 

Pressurization

Pulse: +5%

 

. . . .Manually by inflator bulb

 

Depressurization

. . . .Constant-air release-valve system

Deflation

. . . .Hand-held push button

 

Power source

. . . .1 AA battery (1.5 volt, included)

Battery life

. . . .Approximately 2,000

 

Operating environment

measurements

 

. . . .50ºF to 104ºF (10ºC to 40ºC)

Storage environment .

Less than 85% relative humidity

. . . .-4ºF to 140ºF (-10ºC to 60ºC)

 

Less than 95% relative humidity

Dimensions

. . . .Length: 3.9(100 mm)

 

 

Width: 1.2(31 mm)

 

Weight

Height: 2.0(51 mm)

 

. . . .2.6 oz. (73 g) without battery

Blood pressure measurements determined by the UA-704 are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the ANSI/AAMI SP-10 standard for electronic or automated sphygmomanometers.

In Canada (Toll-Free): 1-800-463-5414

E-15

Page 9
Image 9
A&D UA-704 manual Troubleshooting, Specifications