A&D UA-787 manual Troubleshooting

Models: UA-787

1 27
Download 27 pages 42.31 Kb
Page 24
Image 24

TROUBLESHOOTING

Problem

Probable cause

Corrective action

Nothing appears in

Batteries are drained.

Replace all batteries with new

the display, after I

 

ones. Alkaline batteries

press the START

 

recommended.

button.

 

 

Battery terminals are not in

Reinstall the batteries with

 

 

the correct position.

negative and positive termi-

 

 

nals matching those indicated

 

 

in the battery compartment.

The cuff does not

Battery voltage is too low.

Replace all batteries with new

inflate.

Low battery symbol blinks.

ones. Alkaline batteries rec-

 

[If the batteries are drained

ommended.

 

completely, the mark does

 

 

not appear.]

 

The unit does not

The cuff is not fastened

Fasten the cuff correctly.

properly.

 

measure. Readings

 

are too high or too

You moved your arm or

Make sure you remain very

low.

body during the

still and quiet during the

 

 

measurement.

measurement.

 

The cuff position is not cor-

Sit comfortably and still.

 

rect.

Make sure the cuff is at the

 

 

same level as your heart.

 

You are using the wrong

See Pg. E-5 "Selecting the

 

size cuff.

Correct Cuff."

 

Taking too many readings

Relax five minutes before each

 

on the same arm in a short

measurement.

 

period of time.

 

The value is differ-

The healthcare professional

See Pg. E-18 "Answers to

ent from that

may be using a different

Why Your Readings Are

measured at a clinic

sized cuff.

Different Between Home and

or doctor’s office.

 

the Doctor’s Office."

 

 

 

 

Your measurements may be

See Pg. E-17 "Why Measure

 

elevated by white coat

Blood Pressure at Home."

 

hypertension

 

 

 

 

NOTE: If the actions described above do not solve the problem, call 1-888- 726-9966 (In Canada - 1-800-461-0991). Do not attempt to repair the device yourself.

E-20

LifeSource Health Line (Toll-Free): 1-888-726-9966

Page 24