?WHAT DISPLAY SYMBOLS MEAN

Display

Condition/Cause

Recommended

Symbol

 

Action

 

 

 

 

 

 

 

 

 

 

 

 

The symbol shown while meas-

Measurement is in progress,

 

 

 

 

 

 

 

 

 

 

urement is in progress. It blinks

remain as still as possible.

 

 

 

 

 

while detecting the pulse.

 

 

 

 

 

 

 

 

 

 

 

 

An irregular heartbeat or body

Take measurement again and

 

 

 

 

 

movement may have occurred.

consult with your physician.

 

 

 

 

 

Refer to page E-11 for more infor-

 

 

 

 

 

 

mation on irregular heartbeats.

 

 

 

 

 

 

 

 

 

 

 

 

 

Previous measurement stored in

 

 

 

 

 

 

MEMORY.

 

 

 

 

 

 

 

 

 

 

 

 

 

The battery power indicator dur-

Replace all batteries with new

Battery Full

ing measurement.

ones when the symbol blinks.

 

 

 

 

 

The batteries are low when it

Alkaline batteries recommended.

Battery Low

blinks.

 

 

 

 

 

 

Unstable blood pressure due to

Try the measurement again.

 

 

 

 

 

excessive body movement.

Remain very still during the

 

 

 

 

 

 

measurement.

 

 

 

 

 

 

 

 

 

 

 

 

The systolic and diastolic values

Fasten the cuff correctly, and try

 

 

 

 

 

are within 10 mmHg of each

the measurement again.

 

 

 

 

 

other.

 

 

 

 

 

 

The pressure value did not

Check for air leaks along the

 

 

 

 

 

increase during inflation.

tube and around the

 

 

 

 

 

 

air socket.

 

 

 

 

 

 

 

 

 

 

 

 

The cuff is not fastened correctly.

Refasten the cuff and retake

 

 

 

 

 

 

measurement.

 

 

 

 

 

There is an air leak in the cuff or

Make sure tube is properly con-

 

 

 

 

 

monitor.

nected to cuff and monitor.

 

 

 

 

 

The pulse is not detected correctly.

Try the measurement again.

PULSE DISPLAY ERROR

 

Remain very still during

 

the measurement.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cuff Inflation Meter

Measurement is in progress,

 

 

 

 

 

 

remain as still as possible.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure Rating Indicator™

Refer to the section “About

 

 

 

 

 

 

 

 

 

 

 

Pressure Rating Indicator” for

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

further explanation.

 

 

 

 

 

 

 