Calibration and Test
All ACCES cards are calibrated prior to shipment. However, periodic calibration of
A
Calibration is performed using the SETMUX program supplied with the
Calibration Procedure
The
d.RP1 sets an output offset of the programmable gain amplifier (U13) to zero.
e.RP2 sets an input offset of the programmable gain amplifier (U13) to zero.
f.RP3 sets the level of the +10V regulated supply.
g.RP4 sets the output of the local reference junction temperature sensor.
h.RP5 is a fine gain adjust and provides precise gain setting capability.
The following procedure is brief and is intended for use in conjunction with the calibration part of the SETMUX program.
1.Start the calibration program by typing SETMUX and press the ENTER key at the DOS prompt.
2.Use the relevant menu selection to assign your
3.Use the arrow key to select option 6) Calibrate MUX, then press the ENTER key.
4.Now select which
5.You will be presented with an initial setup for the
6.Use the arrow key to select option 1. Adjust RP3, from the menu at the top left hand corner of the screen.
7.Following the instructions on the screen, perform the RP3 adjustment
8.Repeat steps 6) and 7) for RP4, RP1, RP2, and RP5.
9.This completes the calibration procedure.