Touchscreen (optional)

Table 1.1: Touchscreen Specification

Type

Analog Resistive

 

 

Resolution

Continuous

Light Transmission

75%

Controller

RS-232 interface (uses COM3)

Power Consumption

+5V@200 Ma

Software Driver

Supports Windows 2000,Windows XP

Durability (touches in a lifetime)

30 million

Note! The Point-of-Care Terminal with the optionally installed touchscreen will share COM3. Once the touchscreen is installed, COM3 cannot be used for other purposes.

Optional modules

!Memory: 512 MB, 1 GB, 2 GB DDR2 533/667MHz DRAM

!HDD: 2.5" SATA HDD

!Touchscreen: Analog resistive

Cleaning/Disinfecting

During normal use, the POC-S196 may become soiled and should therefore, be cleaned regularly. Agents: Green tinctured soap and Enzymatic detergents

Steps:

1.Wipe the POC-S196 with a clean cloth that has been moistened in the cleaning solution.

2.Prepare agent per manufacturer’s instructions or hospital protocol.

3.Wipe thoroughly with a clean cloth.

Caution! Do not immerse or rinse the POC-S196 or its peripherals. If you acci- dentally spill liquid on the device, disconnect the unit from the power source. Contact your Biomed department regarding the continued safety of the unit before placing it back in operation.

Do not spray cleaning agent on the chassis.

Do not use disinfectants that contain phenol. Do not autoclave or clean the POC-S196 or its peripherals with strong aromatic, chlorinated, ketone, ether, or Esther solvents, sharp tools or abrasives. Never immerse electrical connectors in water or other liquids.

POC-S196 User Manual