36
Calibration Tests and Adjustment
CMRR Test
CMRR Test
This test checks the CMRR at 1 MHz and 100 MHz.
Specification 3000:1 at 1 MHz, 10:1 at 100 MHz
Equipment Required
Procedure
1Connect the probe power connector to the PROBE connection on the rear of
the 1142A Probe Control and Power Module.
2Set the 1142A front panel switches to Local and Zero offset.
3Connect the probe output to the oscilloscope channel 1 input.
4Connect the input of the probe to the test board in the position shown in the
figure below (signal to + input).
Figure 2-9
Signal to + input
5Connect the signal generator to the test board.
6Set the signal generator for 1 MHz at 385 mVrms (1 Vp-p).
Equipment
Required Critical Specifications
Recommended
Agilent
Model/Part
Oscilloscope 400 MHz bandwidth at 1 mV/div 54830B
Signal Generator 1-100 MHz at = 400 mVrms 8648A
Test Board No substitute 01141-66504
Cable Type N (m) 24-inch 11500B
Adapter Type N (f) to BNC (m) 1250-0077