37
Calibration Tests and Adjustment
CMRR Test
7On the oscilloscope, press AUTOSCALE and set the following parameters.
8On the oscilloscope, measure the peak-to-peak voltage of the channel 1 signal,
then V P-P, then press 1) and record the reading.
Vp-p (1) = _____________ mV
9Connect the input of the probe to the test board in the position shown in the
figure below (signal to both inputs).
Figure 2-10
Signal to both input
10 Set the sensitivity to 1 mV/div.
11 After the measurement settles (averaging is complete), record the V P-P
reading.
Vp-p (2) = _____________ mV
12 Disconnect the probe amp from the test board and measure Vnoise pp on
channel 1.
13 Calculate the CMRR result as follows
14 The result in step 12 should be 3000, representing a CMRR of 3000:1 or more.
Record the CMRR in the Calibration Test Record.
Menu Selection Setting
TIMEBASE (time/div) 500 ns/div
CHAN 1 (sensitivity)
(input R)
200 mV/div
50 DC
ACQUISITION Sampling Mode
Memory Depth
Sample Rate
Averaging
# of avg
Real Time
Automatic
Automatic
Enabled
32
CMRR
Vpp1
Vpp2Vnoisepp
-------------------------------------=