5 KDa MWCO 4 mL |
| 5 KDa MWCO 15 mL | |
30 KDa MWCO 4 mL |
| 30 KDa MWCO 15 mL | |
50 KDa MWCO 4 mL |
| 50 KDa MWCO 15 mL |
Agilent Spin Concentrators for Proteins
Instructions – 4 mL Concentrators
5,000 Molecular Weight
4 mL Spin Concentrator, Part No.
Equipment required | 4 mL Spin Concentrators, pack of 25 |
Centrifuge |
|
Rotor type | Swing bucket or Fixed angle |
Minimum rotor angle | 25° |
Rotor cavity | To fit 15 mL (17 mm) conical bottom tubes |
Minimum effective speed | 2,000 g |
Recommended speed | 5,000 g |
Maximum speed | 7,500 g |
Concentrate recovery |
|
Pipette type | Fixed or variable volume |
Recommended tip | Thin gel loader type |
Technical specifications |
|
Concentrator sample volume capacity |
|
Swing bucket rotor | |
Fixed angle rotor | |
Dimensions |
|
Total Length | 122 mm |
Width | 17 mm |
Active Membrane Area | 2.0 cm2 |
<10 µL | |
Dead stop volume | 20 µL |
pH range | pH 1 – 9 |
Materials of construction |
|
Body | Polycarbonate |
Filtrate vessel | Polypropylene |
Concentrator cap | Polycarbonate |
Membrane | Polyethersulfone |
Concentrating Proteins with 4 mL Concentrators: | Desalting/Buffer Exchange | ||
1 | Fill concentrator with sample within the working volume | 1 | Concentrate sample to desired level or at least |
| capacity range noted in the table above (ensure that the hinged | 2 | 5. |
2 | lid is fully seated). | Empty filtrate container. | |
Insert assembled concentrator into centrifuge (when fixed | 3 | Refill concentrator with an appropriate buffer. | |
| angle rotors are used, angle the concentrators so that the | 4 | Concentrate the sample again and repeat the |
3 | printed windows face upwards/outwards). |
| process until the concentration of the |
Centrifuge at recommended speeds, taking care not to exceed |
| contaminating species is sufficiently reduced. | |
| the maximum g force indicated in the table above (20 minutes |
| Typically three wash cycles will remove 99% of |
| at 7,500 g is recommended for 30 concentration of 4 mL at |
| initial salt content. |
4 | 10C). | See Notes on next page. | |
Once the desired concentration is achieved, remove assembly |
and recover sample from the bottom of the concentrate pocket with a pipette.