40 Language Dictionary
OUTP:REL
Do not install or program the Agilent Relay Accessories if the power supply maximum output current
rating (see Table 3-1) exceeds the contact ratings of the relay.
This command is valid only if the power supply is configured for the optional relay connector. Programming ON closes the
relay contacts; programming OFF opens them. The relay is controlled independently of the output state. If the power
supply is supplying power to a load, that power will appear at the relay contacts during switching. If the power supply is not
configured for the relay option, sending eit her relay command generates an error.
Command Syntax OUTPut:RELay[:STATe] <bool>
Parameters 0 | 1 | OFF | ON
*RST Value 0
Examples OUTP:REL 1 OUTP:REL OFF
Query Syntax OUTPput:RELay?
Returned Parameters 0 |1
Related Commands OUTP[:STAT] *RCL *SAV
OUTP:REL:POL
Do not install or program the Agilent Relay Accessories if the power supply maximum output current
rating (see Table 3-1) exceeds the contact ratings of the relay.
This command is valid only if the power supply is configured for the optional relay connector. Programming NORMal
causes the relay output polarity to be the same as the power supply output. Programming REVerse causes the relay output
polarity to be opposite to that of the power supply output. If OUTP[:STAT] = ON when either relay command is sent, the
power supply output voltage is set to 0 during the time that the relays are changing polarity.
If the power supply is not configured for the relay option, sending either relay command generates an error.
Command Syntax OUTPut:RELay:POLarity <CRD>
Parameters NORMal | REVerse
*RST Value NORM
Examples OUTP:REL:POL NORM
Query Syntax OUTPput:RELay:POLarity?
Returned Parameters NORM | REV
Related Commands OUTP[:STAT] *RCL *SAV