1-4
Installation
Installation
Figure1-2. Position of modules in the mainframe
3Clean all optical interfaces as described in “Cleaning Connections for Accurate
Measurements” on page 4-10, before making measurements.
4Perform the following steps if you’re installing an Agilent 83492/3A module:
aUnscrew and remove the fiber-optic adapter that is located on the optical
module’s front-panel optical input connector.
bConnect the adapter that was removed in the previous step onto the
Agilent 83492/3A module’s front-panel Input connector.
On Agilent 83492A module’s, the input connector used depends on the
wavelength of the input signal. Refer to “To connect cables to an
Agilent 83492A” on page 1-7.
5Connect the supplied adapter cable as shown in Figure 1-3.
WARNING Light energy can radiate from the front pane l OUTPUT connectors on
Agilent 83492A and 83493A modules. The light emitted from these
connectors is the slightly attenuated light that is i nput to the front-
panel INPUT connector.
6Turn on the Agilent 83480A, and connect a modulated signal to the
Agilent 83491/2/3A module’s Input connector.