These specifications apply to the Agilent
Technologies 8591E, 8593E, 8594E, 8595E, and
8596E spectrum analyzers.

Specifications

All specifications apply over O°C to +55°C. The
analyzer will meet its specifications after 2 hours
of storage at a constant temperature, wit hin the
operating temperature range, 30 minutes after t he
analyzer is tumed on, and after CAL FREQ and
CAL AMPTD (and for the 8593E, 8595E, and
8596E CAL YTF) have been run.

Frequency Specifications

Frequency Range
8591E
50 9 kHz to 1.8 GHz
75 1 MHz to 1.8 GHz
8593E 9 kHz to 22 GHz
Option 026/027 9 kHz to 26.5 GHz
Band LO harmonic = N
0 1 9 kHz to 2.9 GHz
1 1 2.75 GHz to 6.5 GHz
2 2 6.0 GHz to 12.8 GHz
3 3 12.4 GHz to 19.4 GHz
4 4 19.1 GHz to 22.0 GHz
4 4 (Opt. 026/027) 19.1 GHz to 26.5 GHz
8594E
dc coupled 9 kHz to 2.9 GHz
ac coupled 100 kHz to 2.9 GHz
8595E
dc coupled 9 kHz to 6.5 GHz
ac coupled 100 kHz to 6.5 GHz
8596E
dc coupled 9 kHz to 12.8 GHz
ac coupled 100 kHz to 12.8 GHz
Band LO harmonic = N
0 1 9 kHz to 2.9 GHz (dc coupled)
0 1 100 kHz to 2.9 GHz (ac coupled)
1 1 2.75 GHz to 6.5 GHz
2 2 6.0 GHz to 12.8 GHz
Frequency Reference (Opt. 004)
Aging ±2 x 10-6/year ±1 x 10-7/year
Temperature Stability ±5 x 10-6 ±1 x 10-8
Initial Achievable Accuracy ±0.5 x 10-6 ±2.2 x 10-8
Agilent 8590 E-SeriesPortable Spectrum Analyzers

Data Sheet