3
Amplitude Specifications

Amplitude specifications do not apply for Analog+

mode and negative peak detector mode except as

noted in “Amplitude Characteristics.”

Amplitude Range

Displayed average noise level
to +30 dBm
8591E (Opt. 001) Displayed average noise level to
+72 dBmV
Maximum Safe Input Level (input attenuator 10 dB)
Average Continuous Power +30 dBm (1 W)
8591E (Opt. 001) +72 dBmV (0.2 W)
Peak Pulse Power
8591E +30 dBm (1 W)
8591E (Opt. 001) +72 dBmV (0.2 W)
8593E, 94E, 95E, 96E +50 dBm (100 W) for < 10 µs pulse
width and <1 % duty cycle, input
attenuation 30 dB.
dc
8591E 25 Vdc
8591E (Opt. 001) 100 Vdc
8593E 0 Vdc
8594E, 95E, 96E 0 V (dc coupled)
50 V (ac coupled)

Gain Compression

>10 MHz 0.5 dB (total power at input mixer2 = –10 dBm)

Displayed Average Noise Level

(Input terminated, 0 dB attenuation, 1 Hz/30 Hz VBW, sample-
detector)
8591E 30 Hz RBW 1 kHz RBW
400 kHz to 1 MHz –130 dBm –115 dBm
1 MHz to 1.5 GHz –130 dBm –115 dBm
1.5 GHz to 1.8 GHz –128 dBm –113 dBm
8591E (Opt. 001)
1 MHz to 1.5 GHz –78 dBmV –63 dBmV
1.5 GHz to 1.8 GHz –76 dBmV –61 dBmV
8593E
400 kHz to 2.9 GHz –127 dBm –112 dBm
2.75 GHz to 6.5 GHz –129 dBm –114 dBm
6.0 GHz to 12.8 GHz –117 dBm –102 dBm
12.4 GHz to 19.4 GHz –113 dBm –98 dBm
19.1 GHz to 22 GHz –107 dBm –92 dBm
8593E (Opt. 026/027)
19.1 GHz to 26.5 GHz –102 dBm –87 dBm
8594E
400 kHz to<5 MHz –122 dBm –107 dBm
5 MHz to 2.9 GHz –127 dBm –112 dBm
8595E
400 kHz to 2.9 GHz –125 dBm –110 dBm
2.75 GHz to 6.5 GHz –127 dBm –112 dBm
8596E
400 kHz to 2.9 GHz –125 dBm –110 dBm
2.75 GHz to 6.5 GHz –127 dBm –112 dBm
6.0 GHz to 12.8 GHz –115 dBm –100 dBm
Sensitivity 30 Hz RBWThird Order IntermodSecond Order Distortion30405060708090100–70 –60 –50 –40 –30 –20 –10 0 +10
19.1 TO 22 GHz
6.0 TO 12.8 GHz
400 kHz TO 6.5 GHz
12.4 TO 19.4 GHz
>2.75 GHz

Nominal Dynamic Range

Agilent 8593E (Option 130)

Mixer Level (dBm)

Dynamic Range (dB)

(30 Hz Res BW)
10 MHz to 2.9 GHz
2. Mixer Power Level (dBm) = Input Power (dBm) Input Atten. (dB)