1-27
Setting Up the Infiniium DCA
Step 12. Perform a Module Calibration
To perform a module calibration
1On the Calibrate menu, choose All Calibrations.
The All Calibrations dialog box opens. This dialog box allows you to view
the calibration status of modules and initiate the calibration.
2Touch the Modules tab.
The Modules tab opens and allows you to perform a module calibration on
either the left or right module.
3Touch the Calibrate Module button (located at the bottom of
the tab) for the module you need to c alibrate. This starts the
Module calibration.
4Remove all external devices and signals from the module.
The module calibration uses known s ignal levels in the instrument. Introduc-
ing an outside signal source will interfere with the calibration factors and
decrease the accuracy of the calibration.
5Touch Continue. A progress indicator appears.
6After the calibration completes, choose All Calibrations on the Calibration
menu.
7The Module tab shows the status of the calibration. The tab will look similar to
the following figure. Notice that the tab indicates tha t the “Current user