Page 2 | Service Note |
Solution/Action:
Configure the unit to output a ramp waveform from OUTPUT 1 at approximately 1 kHz and 1 Volt. Verify with an oscilloscope whether the ramp waveform is inverted (ramps down instead of up). If this behavior is unwanted then replace T1 with the part number noted above.