1-10
Setting Up the Infiniium DCA
Step 3. Consider Environmental Specifications
Step 3. Consider Environmental Specifications

Review the following specifications to ensure that your operating or storage

environment is suitable for the instrument.

Table1-6. Environmental Specifications

Use indoor
Temperature
Operating 10 °C to +40 °C (50 °F to + 104 °F)
Non-operating -40 °C to +70 °C (-40 °F to +158 °F)
Humidity
Operating Up to 90% humidity (non-condensing) at +40 °C (+104 °F)
Non-operating Up to 95% relative humidity at + 65 °C (+149 °F)
Altitude
Operating Up to 4,600 meters (15,000 ft)
Non-operating Up to 15,300 meters (50,000 ft)
Vibration
Operating Random vibration 5-500 Hz, 10 minutes per axis, 2.41 g(rms)
Non-operating Random vibration 5-500 Hz, 10 minutes per axis, 0.3 g(rms);
Resonant search, 5-500Hz swept sine, 1 octave/minute sweep
rate, 0.75g, 5 minute resonant dwell at 4 resonances per axis.
Power requirements
Voltage 90 to 132 or 198 to 264 Vac, 48-66 Hz CAT II
Power (including modules) 545 VA; 375 W
Weight
Mainframe 15.5 kg (34 lb)
Modules 1.2 kg (2.6 lb)
Dimensions
Height 215.1 mm (8.47 in)
Width 425.5 mm (16.75 i n)
Depth 566 mm (22 .2 in)
Depth with front connectors
and rear feet
629 mm (24.8 in)