87
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
logic analyzer is still running.
Messages such as "Waiting in level 1" may indicate you need to refine your
trigger.
If the status shows as "Stopped", the analyzer either finished the
acquisition, or was unable to run. T he cause of the problem is listed in the
bottom half of the Run Status window.
Look for an error message in the message bar at the top o f the window.
Common messages are "slow or miss ing clock" and "Waiting for trigger".
If Run briefly changed to Stop or Cancel, select the Window menu,
choose the logic analyzer's slot, then choose the Waveform or Listing
display.
See Also “Slow or Missing Clock” on page 180
“Waiting for Trigger” on page 184
“Error Messages” on page 170
To view the trigger status

While a logic analyzer measurement is running, you can view the

trigger status to see the sequence level that is evaluating captured

data, occurrence and global counter values, and flag values.

1. In the Trigger tab, select the Status subtab.
See Also Run status (in the system help volume).